摘要 目的建立化妆品中六价铬的离子色谱-紫外检测器测定方法。方法样品经0.05 mol/L氢氧化钠-磷酸盐缓冲液超声提取后离心,经阴离子色谱柱IonPac AS7(4 mm×250 mm)分离,采用150 mmol/L氯化铵-60 mmol/L氨水溶液为淋洗液,流速为0.8 ml/min,以紫外检测器(375 nm)检测。结果该方法线性范围为0.005~1 mg/L,回归方程为y=18.923x-0.011 7,r>0.999 9。方法检出限为0.04 mg/kg,定量下限为0.12 mg/kg,平均回收率为88.2%~98.8%,RSD为2.24%~4.50%。结论该方法操作简便,具有较好的精密度和准确度,适用于化妆品中六价铬的检测。 Objective To establish an analytical method for the determination of hexavalent chromium in cosmetics by ion chromatography with UV detector. Methods The cosmetics samples were treated by ultrasonic extraction with 0.05 mol/L sodium hydroxide +phosphate buffer solution and then centrifuged,hexavalent chromium in extract were separated by anion chromatography column IonPac AS7(4 mm ×250 mm),with mobile phase of 150 mmol/L ammonium chloride-60 mmol/L ammonia solution,the flow rate was 0.8 ml/min,The UV detection wavelength was 375 nm. Results A good linear relationship showed in the range of 0.005-1 mg/L,and the regression equation was y=18.923 x-0.011 7,r>0.999 9. The limit of detection was 0.04 mg/kg,the limit of quantification was 0.12 mg/kg,the average recovery rates were 88.2%-98.8%,RSDs were 2.24%-4.50%. Conclusion This method is simple,accurate,precise and applicable to the determination of hexavalent chromium in cosmetics.
作者 滕小沛 沈宗岚 张立辉 TENG Xiao-pei;SHEN Zong-lan;ZHANG Li-hui(Institute for Physical and Chemical Test ,Jiangsu Provincial Center for Disease Control and Prevention, Nanjing ,Jiangsu 210009, China)
出处 《环境与健康杂志》 CAS 北大核心 2018年第12期1072-1074,共3页 Journal of Environment and Health
基金 江苏省医学重点学科(实验室)项目(ZDXKA2016008)
关键词 六价铬 化妆品 离子色谱-紫外检测 Hexavalent chromium Cosmetics Ion chromatography-UV detection
  • 相关文献

参考文献5

二级参考文献90

  • 1黄晖,陈培珍.浊点萃取技术在痕量金属离子测定中的应用[J].三明高等专科学校学报,2004,21(4):66-71. 被引量:12
  • 2王柏松,李华,丁富新,万俊生,王武尚.毛细管电泳柱上络合分离、间接紫外检测重金属离子[J].分析测试学报,2006,25(5):25-29. 被引量:3
  • 3罗武庭.DJ—2可变矩形电子束曝光机的DMA驱动程序[J].LSI制造与测试,1989,10(4):20-26. 被引量:373
  • 4TANG An -na,JIANG Dong-qing,JIANG Yan,et al.Cloud point extraction for high-performance liquid chromatographic speciation of Cr(Ⅲ) and Cr(Ⅵ) in aqueous solutions[J].Journal of Chromatography A,2004,1036(2):183-188.
  • 5BOBROWSKI A,KRLOICKA A,ZAR E,BSKI J.Characteristics of voltammetric determination and spociation of chromium-a review[J].Electroanalysis,2009,21 (13):1449-1458.
  • 6CANDIR S,NARI I,SOYLAK M.Ligandless cloud point extraction of Cr(Ⅲ),Pb(Ⅱ),Cu(Ⅱ),Ni(Ⅱ),Bi(Ⅲ),and Cd(Ⅱ) ions in environmental,samples with Tween 80 and flame atomic absorption spectrometric determination[J].Talanta,2008,77:289-293.
  • 7FLORAS De. Threshold mechanisms and site specificity in chromium(Ⅵ) carcinogenesis[ J]. Carcinogenesis, 2000, 21 (4) : 533 -541.
  • 8FANG Guozhen, LUO Jikuen. Spectrophotometric determination of chromium ( Ⅲ, Ⅵ ) by use of chromium ( Ⅲ, Ⅵ ) chmm A zurol S-cetylpyridinium bromide-hydroxylamine hydrochloride-zinc ( Ⅱ ) system [ J ]. Talanta, 1992, 39 (12) : 1579 - 1582.
  • 9SCINDIA Y M, PANDEY A K, REDDY A V, et al. Selective preconcentration and determination of chromium(Ⅵ) using a flat sheet polymer inclusion sorbent : potential application for Cr ( Ⅵ ) determination in real samples [ J ]. Anal Chem, 2002, 74(16) : 4204 -4212.
  • 10SAMANTA G, BORING C B, DASGUPTA P K. Continuous automated measurement of hexavalent chromium in airborne particulate matter[ J ]. Anal Chem, 2001, 73 (9) : 2034 - 2040.