期刊文献+

实时三维超声心动图对经导管闭合房间隔缺损术后的观察

Application of real-time three-dimensional echocardiography to the observation of percutaneous transcatheter closure of atrial septal defect
暂未订购
导出
摘要 目的 :探讨实时三维超声心动图 (RT- 3DE)对经导管闭合房间隔缺损 (ASD)术后观察的价值。方法 :对 15例ASD患者术后封堵器的位置、形态、塑型、大小及运动进行 RT- 3DE观察 ,并与手术结果进行比较。结果 :RT- 3DE显示封堵器呈双盘状 ,测得封堵器伞腰直径为 2 7± 5 mm ,与实际封堵器大小 (2 6± 5 mm )比较无显著差异 (P>0 .0 5 ) ;封堵器相对房间隔无移位 ,随心脏运动 12例无扩展和收缩 ,形状也无变化 ,3例有变化。结论 :RT- 3DE成像速度快 ,模拟外科手术路径实时显示封堵器的位置、形态和塑形以及和周围组织的关系 ,可作为 ASD介入治疗术后疗效评价的重要手段。 AIM: To evaluate the application of real-time three-dimensional echocardiography (RT-3DE) to the observation of curative effect of interventional treatment of Atrial septal defect (ASD) with Amplatzer closure devices. METHODS: Sizes, shapes, sites and movements of devices after the treatment were observed with 3DE in 15 secudum ASD patients. RESULTS: RT-3DE could clearly visualize the device's double-disc like shapes and sites in 3 dimensions. There was no significant difference between the measurements of occluder waist sizes with 3DE and actual ones (27±5mm vs 26±5 mm, P>0.05). 12 devices did not move in cardiac cycles and their shapes did not change except those of 3 cases. CONCLUSION : RT-3DE is a high frame rate imaging technology, which can provide clear real-time illustration of shapes, sites, remodeling and motion of Amplatzer closure devices by simulating an operation. It can be used as an important method for evaluating interventional treatment of ASD.
出处 《心脏杂志》 CAS 2004年第3期270-272,共3页 Chinese Heart Journal
关键词 超声心动图 实时 三维 房间隔缺损 封堵术后 echocardiography, real-time,three-dimension atrial septal defect percutaneous transcatheter closure
  • 相关文献

参考文献11

二级参考文献48

  • 1李国业,熊遂秀,李其果.房间隔穿刺定位法[J].临床心血管病杂志,1989,5(1):24-26. 被引量:18
  • 2张显利,应大君,姚远,李震一,鲁云敏.气囊二尖瓣成形术的应用解剖[J].中国临床解剖学杂志,1993,11(3):209-210. 被引量:3
  • 3纪荣明,党瑞山,姜宗来,何北平,秦永文.卵圆窝穿刺定位的应用解剖[J].中国临床解剖学杂志,1995,13(1):36-37. 被引量:2
  • 4[1]King TD, Mills NL. Nonoperative closure of atrial septal defects. Surgery, 1974,75:383-388.
  • 5[2]Rao PS, Wilson AD, Levy JM, et al. Role of “buttoned” double-disc device in the management of atrial septal defects. Am Heart J, 1992,123:191.
  • 6[3]Masura J, Gavora P, Formanek A, et al. Transcatheter closure of secundum atrial septal defects using the new self-centering amplatzer septal occluder: initial human experience. Cathet Caridiovasc Diagn,1997,42:388.
  • 7[4]Thanopoulos BD, Laskari CV, Tsaousis GS, et al. Closure of atrial septal defects with the Amplatzer occlusion device: preliminary results. J Am Coll Cardiol,1998,31:1110.
  • 8[5]Zamora R, Rao PS, Lloyd TR, et al. Intermediate-term results of phase I Food and Drug Administration Trials of buttoned device occlusion of secundum atrial septal defects. J Am Cool Cardiol,1988,31:674-676.
  • 9[6]Hijazi ZM, Bova MPHS, Rhodes J. Simultaneous transcatheter cloure of two secundum atrial septal defect using the AmplatzerTM septal occluder. J Intervent Cardiol,1998,11:181.
  • 10[7]Sharafuddin MJ, Gu X, Titus JL, et al. Transvenous closure of secundum atrial septal defects: preliminary results with a new self-expanding nitinol prosthesis in a swine model. Circulation,1997,95:2162.

共引文献171