期刊文献+

激光量热法测量红外光学薄膜吸收 被引量:2

Laser calorimeter for IR absorption measurement of thin films
在线阅读 下载PDF
导出
摘要 建立了一种评价波长为1.319μm光学薄膜的吸收损耗的吸收测量系统。采用激光量热计方法的原因在于:它已被广泛用来测量光学薄膜吸收,并且有很高的可靠性。描述了激光量热计法测量吸收装置的设计和建立,并给出基底—薄膜整体吸收的测量结果和记录的温度曲线,最后提出了在研究激光量热计测量方法中几个要考虑的问题。 An absorptance measurement system has been developed for evaluation of the absorption loss of optical coatings at λ=1.319μm(YAG laser).Calorimetry was adopted as the measurement method because of its widely used to measure thin film optical absorption and its high reliability.The design and construction of laser calorimeter have been discribed. The result of the total film-substrate system absorptance and the recorded temperature-curve have been presented in this paper. At the same time ,Serval questions of the calorimetric measurement have been presented 
出处 《光学仪器》 2004年第2期194-197,共4页 Optical Instruments
关键词 红外光学薄膜 吸收比 激光量热法 测量 激光量热计 absorptance calorimetry YAG laser thin film
  • 相关文献

参考文献5

  • 1丁丽娟.数值计算方法[M].北京:北京理工大学出版社,2000..
  • 2ISO11551-1997, Optics and optical instruments-lasers and laser-related equipment-test method for absorptance of optical laser components[S].
  • 3Chidane O,Masanobu H,Akin M,Kazuho S.Measurement of absorptance of optical coatings for F2 lithographe[J].SPIE, 2001,4449 :7-12.
  • 4Atkinson R.Development of a wavelength scanning laser calorimeter[J].Applied Optics,1985,24(4/15):464-471.
  • 5Marvin H,James W D, Herbert B R,Julius B.Measurement of very low absorption coefficients by laser calorimetry[J].Applied Optics,1975,14(5):1128-1130.