期刊文献+

源错位对冠状动脉内近距离放射治疗剂量的影响 被引量:1

Effect on Dose Owing to Source Displacement in Coronary Endovascular Brachytherapy
暂未订购
导出
摘要 目的 研究以导管为基础192Ir放射性线源治疗冠状动脉再狭窄时,放射源与靶区错位对靶区边缘剂量的影响。方法 用AAPM60,43号报告推荐的公式计算剂量分布。结果 放射源与靶区错位1.1 mm时,靶区边缘剂量最大可降低53.9%,受影响范围为6 mm;放射源与靶区错位5 mm时,靶区边缘剂量最大可降低565.8%,受影响范围为9 mm。结论 错位将导致一端靶区边缘的剂量降低。 Objective To investigate the effect on the dose distribution of source position displacement with the target region margin in catheter-based 192Ir line source endovascular brachytherapy. Method Dose rate distribution along longitudinal axes was estimated by the formula recommended by AAPM No.60 and No.43. Results In the two cases of source displacement (1.1 and 5 mm) doses of target region margin was decreased dramatically (at most 53.9% and 565.8% respectively) were compared to that of no source displacement, and the affected range was 6 mm and 9 mm respectively. Conclusion Source displacement will lead to the decrease of dose in target region margin.
机构地区 中国医学科学院
出处 《中国医学科学院学报》 CAS CSCD 北大核心 2003年第2期193-196,共4页 Acta Academiae Medicinae Sinicae
基金 国家自然科学基金(30170233)资助
关键词 再狭窄 边缘效应 以导管为基础的冠状动脉近距离治疗 源错位 restenosis edge effect catheter-based endovascular brachytherapy source displacement
  • 相关文献

参考文献7

  • 1徐志勇,张良安,戴光复,王运来.用蒙特卡罗方法估算^(60)Co辐射源事故患者的辐射剂量[J].中华放射医学与防护杂志,2001,21(3):159-161. 被引量:7
  • 2徐志勇,张良安,戴光复.蒙特卡罗法和积分法计算核素治疗后的病人对周围人员的照射量[J].中华放射医学与防护杂志,2001,21(4):253-254. 被引量:7
  • 3Giap HB, Bendre DD, Huppe GB, et al. Source displacement during the cardiac cycle in coronary endovascular brachytherapy. Int J Radiat Oncol Biol Phys, 2001, 49:273-277.
  • 4Nath R, Amols H, Coffey C, et al. Intravascular brachytherapy physics: report of the AAPM radiation therapy committee task Group No. 60. American Association of Physicists in Medicine. Med Phys, 1999, 26:119-152.
  • 5Teirstein PS, Kuntz RE. New frontiers in interventional cardiology: intravascular radiation to prevent restenosis. Circulation, 2001, 104:2620-2626.
  • 6Kim HS, Waksman R, Cottin Y, et al. Edge stenosis and geographical miss following intracoronary gamma radiation therapy for in-stent restenosis. J Am Coil Cardiol, 2001, 15(37): 1026-1030.
  • 7Nath R, Anderson L, Luxton G, et al. Dosimetry of interstitial brachytherapy sources: recommendations of the AAPM Radiation Therapy Committee Task Group No.43. Med Phys,.1995, 22:209-234.

二级参考文献7

  • 1许淑艳.蒙特卡罗方法在实验物理中的应用[M].北京:原子能出版社,1995.5-19.
  • 2Cho S H,Int J Radiat Oncol Biol Phys,1997年,37卷,483页
  • 3许淑艳,蒙特卡罗方法在实验核物理中的应用,1995年,5页
  • 4李士骏,电离辐射剂量学(第2版),1986年,215页
  • 5Snyder W S,MIRD Pamphlet.5,1978年
  • 6Snyder W S,J Nucl Med,1969年,10卷,supp 3期,5页
  • 7Dimos B,Stavroula G,Anastasia,et al.Application of the Monte Carlo integration method for calculation of the anisotropy of 192 Ir brachytherapy sources[].Physics in Medicine and Biology.1998

共引文献7

同被引文献23

  • 1Tung C J,Liu C S,Wang J P,et al.Calculations of Cellular Microdosimetry Parameters for Alpha Particles and Electrons.Applied Radiation and Isotopes,2004,61:739.
  • 2Cruza G S,Palmer M R,Matatagui E,et al.A Theoretical Model for Event Statistics in Microdosimetry.Ⅱ: Nonuniform Distribution of Heavy Ion Tracks.Medical Physics,2001,28(6):997.
  • 3Sankaranarayanan V,Ganesan S,Oommen S,et al.Study on Dosimetric Parameters for Stereotactic Radiosurgery and Intensity-modulated Radiotherapy.Med.Dosim.,2003,28(2):85.
  • 4Kirisits C,Hefner A,Wexberg P,et al.Estimation of Doses to Personnel and Patients During Endovascular Brachytherapy Applications.Radiat.Prot.Dosimetry.2004,108(3):237.
  • 5Chen F M,Epstein A L,Li Z,et al.Comparative Autoradiographic Study Demonstrating Differential Intratumor Localization of Monoclonal Antibodies to Cell Surface(Lym-1)and Intracellular (TNT-1)Antigens.J.Nucl.Med.,1990,31(9):1059.
  • 6Gardin I,Linhart N C,Petiet A,et al.Dosimetry at the Cellular Level of Kupffer Cells after Technetium-99m-sulphur Colloid Injection.J.Nucl.Med.,1992,33(3):380.
  • 7Puncher M R,Blower P J.Frozen Section Microautoradiography in the Study of Radionuclide Targeting: Application to Indium-Ⅲ-Oxine-Labded Leukocytes.J.Nucl.Med.,1995,36(4):499.
  • 8Goddu S M,Howell R W,Bouchet L G,et al.MIRD Cellular S Values.Society of Nuclear Medicine,850 Samuel Morse Drive,Reston,VA 20190~5136,1997.
  • 9Laniece A,Pihet P,Winkler R,et al.Monte Carlo Code for Microdosimetry of Inhaled α Emitters.Radiat.Prot.Dosimetry,2002,99:463.
  • 10Stewart R D,Wilson W E,Mcdonald J C,et al.Microdosimetric Properities of Ionizing Electrons in Water: a Test of the PENELOPE Code System.Phys.Med.Biol.,2002,47:79.