摘要 目的探讨POCT便携式血糖仪与全自动生化分析仪检测比对结果,定期评价便携式血糖仪的可靠性。方法选取22位空腹血糖在0.88~30.82mmol/L的患者血样,置于氟化钠采血管中,分别用7600全自动分析系统与POCT便携式血糖仪检测,每份样本测定3次,取均值,并对结果进行偏倚及相关性评价。结果20台血糖仪偏倚在±0.2以内,PT得分大于80%,5台仪器分别有2~6个浓度的测定值偏倚超出±0.2。结论 20台便携式血糖仪检测结果与全自动生化分析仪检测结果一致,达到卫生部规定的准确性标准,5台血糖仪差异较大。 Objective To compare the results of blood glucose detected by the POCT glucose meters and the automatic biochem‐ical analyzer ,and to periodically evaluate the reliability of the portable glucose meters .Methods 22 patients with fasting blood‐glu‐cose of 0 .88 - 30 .82 mmol/L were selected and the NaF anticoagulation whole blood was collected .Each sample were detected for 3 times .The mean value was taken for conducting the bias and correlation evaluation .Results The bias of 20 glucose meters was within ± 0 .2 and PT 〉 80% ,and the bias of 5 glucose meters ±0 .2 in 2 - 6 concentration values .Conclusion The testing results of 20 glucose meters are identical with those detected by the biochemical analyzer and meet the accuracy standards stipulated by the Ministry of Public Health ,and 5 glucose meters have relatively big differences .
作者 张静
出处 《国际检验医学杂志》 CAS 2015年第6期734-735,737,共3页 International Journal of Laboratory Medicine
基金 内蒙古自然科学基金资助项目(2013MS1116)
关键词 血糖仪 血糖 全自动生化分析仪 glucose meter blood glucose automatic biochemistry analyzer
  • 相关文献

参考文献6

  • 1Hasslacher C,Kulozik F,Platten I. Analytical performance of glucose monitoring systems at different blood glucose ranges and a-nalysis of outliers in a clinical setting[J]. J Diabetes Sci Technol,2014,8(3):466-472.
  • 2Junker R,Schlebusch H, Luppa PB. Point-of-care testing in hospi-tals and primary care[J3- Dtsch Arztebl Int, 2010,107 (33) ! 561-567.
  • 3Kos S, van Meerkerk A,van der Linden J, et al. Validation of anew generation POCT glucose device with emphasis on aspectsimportant for glycemic control in the hospital care[J]. Clin ChemLab Med,2012,50(9) ;1573-1580.
  • 4康熙雄.床旁即刻临床检测的规范化趋向[C].中华医学会第七次全国检验医学学术会议,2008:76-78.
  • 5樊晓萍,张环生,赵秀风,王国萍.快速血糖仪与全自动生化分析仪检测血葡萄糖结果对比研究[J].国际检验医学杂志,2006,27(3):287-287. 被引量:54
  • 6Louie RF, Ferguson WJ,Sumner SL, et al. Effects of dynamictemperature and humidity stresses on point-of-care glucose tes-ting for disaster care[j]. Disaster Med Public Health Prep,2012,6(3):232-240.

二级参考文献4

  • 1Carr SR,Slocum J,Tefft L,et al.Precision of office-based blood glucose meters in screening for gestational diabetes.Am J Obstet Gynecol,1995,173(4):1267-1272.
  • 2Chen HS,Kuo BI,Hwu CM,et al.Technical and clinical evaluation of an electrochemistry glucose meter:experience in a diabetes center.Diabetes Res Clin Pract,1998,42(1):9-15.
  • 3Johnson RN,Baker JR.Accuracy of devices used for self-monitoring of blood glucose.Ann Clin Biochem,1998,35 (Pt 1):68-74.
  • 4张文艳,王明山,刘勇,赵秀琴.血糖监测仪测定微血管全血糖的临床研究[J].天津医科大学学报,1999,5(3):46-46. 被引量:24

共引文献56

同被引文献65

引证文献8

二级引证文献36