期刊文献+

六西格玛度量评价Beckman Coulter LH750血细胞分析仪性能 被引量:9

Six Sigma metrics for evaluating performance of Beckman Coulter LH750 blood cell analyzer
暂未订购
导出
摘要 目的:评价Beckman Coulter L H750检测系统的性能。方法根据CLSI EP9确定分析仪的真实度,根据CLSI EP5确定分析仪的精密度,通过六西格玛度量方法计算血细胞分析仪各项目的性能并进行评估。结果 L H750白细胞计数(WBC )的六西格玛水平在5.5~11.1,红细胞计数(RBC)在5.5~9.4,血红蛋白含量(Hb)在6.0~14.6,血小板计数(PLT )在4.1~10.0,红细胞压积(HCT)的六西格玛值低于了3.0。结论 Beckman Coulter LH750检测系统基本满足六西格玛的质量管理要求, HCT这个项目还需要进行质量改进。六西格玛质量理论可以反映检测过程的性能。 Objective To evaluate the performance of Beckman Coulter LH750 detection system .Methods The truth degree of the analyzer was detected according to the CLSI EP9 and the precision was determined according to the CLSI EP5 analyzer ,the per-formance of various items in the blood cell analyzer was calculated and evaluated by the Six Sigma metrics .Results The Six Sigma level of white blood cell (WBC) count in LH750 was 5 .5-11 .1 ,which of the red blood cell (RBC) count in 5 .5-9 .4 ,which of he-moglobin (Hb) in 6 .0-14 .6 and which of the platelet (PLT) count in 4 .1-10 .0 ,the Six Sigma value of hematocrit (HCT) was lower than 3 .0 .Conclusion The Beckman Coulter LH750 detection system basically meets the requirements of the Six Sigma quali-ty management ,the HCT item also needs to be improved in the quality .The Six Sigma quality theory can reflect the performance of the detection process .
作者 王双 邱世飏
出处 《国际检验医学杂志》 CAS 2014年第5期605-606,共2页 International Journal of Laboratory Medicine
关键词 六西格玛 偏倚 实验室技术和方法 six sigma bias laboratory techniques and procedures
  • 相关文献

参考文献4

二级参考文献17

  • 1莫凡,黄与双,赵洁,肖桂初.6σ质量标准在血细胞分析中的应用研究[J].国际检验医学杂志,2007,28(1):14-16. 被引量:4
  • 2李园园,李萍,黄亨建.应用六西格玛理论评价临床实验室检验性能及设计质控方案[J].中国医疗器械信息,2007,13(6):9-11. 被引量:36
  • 3Westgard JO. Six sigma table EOL/S]. http://www.westgard.con./sixsigmatable.html.2007.
  • 4刘忠民 高月亭 陈涛 等.应用操作过程规范图评价和设计室内质量控制方法.医学实验室与临床,2007,14(4):17-20.
  • 5Nevalainen D,Berte L,Kraft C,et al.Evaluating laboratory performance on quality indicators with the six sigma scale[J].Arch Pathol Lab Med,2000,124(4),516-519.
  • 6Westgard JO.Six Sigma Quality Design and Control[M].2nd ed.Madison:Westgard QC2006:157-184.
  • 7Westgard JO,Westgard SA.The quality of laboratory testing today:an assessment of σ metrics for analytic quality using performance data from proficiency testing surveys and the CLIA criteria for accep table performance[J].Am J Clin Pathol,2006,125(3):343-354.
  • 8Coskun A.Six sand calculated laboratory tests[J].Clin Chem,2006,52(4):770-771.
  • 9Alexander Bragat.Westgard QC A Six Sigma Method Decision Chart spreadsheet.www.westgard.com.
  • 10Westgard JO,Westgard SA.Equivalent quality testing versus equivalent QC procedures[J].Lab Medicine,2005,36(10):626-629.

共引文献38

同被引文献54

引证文献9

二级引证文献58

相关作者