期刊文献+

罗氏血凝仪CoaguChek XS在房颤抗凝管理体系中的应用

原文传递
导出
摘要 目的探讨CoaguChek XS血凝仪在房颤抗凝管理体系中的应用效果。方法对62例口服华法林的房颤患者分别应用血凝仪末梢血检测法和实验室静脉全血检测法检测国际标准化比值(INR),评价CoaguChek XS血凝仪测定INR值的准确性。结果62例患者末梢血检测INR平均值为(2.04±0.57),静脉全血INR平均值为(2.06±0.56),两种方法比较差异无统计学意义(t=0.911,P=0.366)。结论CoaguChek XS血凝仪有望在房颤抗凝管理体系中推广应用。
出处 《中华现代护理杂志》 2013年第6期693-694,共2页 Chinese Journal of Modern Nursing
  • 相关文献

参考文献10

  • 1Camm AJ, Kirchhof P, Lip GY, et al. Guidelines for the management of atrial fibrillation: the Task Force for the Management of Atrial Fibrillation of the European Society of Cardiology (ESC). Europace,2010,12(10) : 1360-1420.
  • 2王晓俏,陈样新.华法令抗凝治疗监测的相关问题探讨[J].临床内科杂志,2008,25(1):50-51. 被引量:8
  • 3Wieloch M, Hillarp A, Strandberg K, et al. Comparison and evaluation of a point-of-care device ( CoaguChek XS) to Owren- type prothrombin time assay for monitoring of oral anticoagulant therapy with warfarin. Thromb Res,2009,124 ( 3 ) :344-348.
  • 4Sobieraj-Teague M, Daniel D, Farrelly B, et al. Accuracy and clinical usefulness of the CoaguChek-S and XS point of care devices when starting warfarin in ahospital outreach setting. Thromb Res ,2009 ,123 ( 6 ) :909-913.
  • 5Plesch W, Wolf T, Breitenbeek N, et al. Results of the performance verification of the CoaguChek XS system. Thromb Res,2008,123 (2) :381-389.
  • 6Nam MH, Roh KH, Rak HN, et al. Evaluation of the Roche CoaguChek XS handheld coagulation analyzer in a cardiac outpatient clinic. Ann Clin Lab Sci,2008,38( 1 ) :37-40.
  • 7Karon BS, Mc Bane RD, Chaudhry R, et al. Accuracy of capillary whole blood intemational normalized ratio on the CoaguChek S, CoaguChek XS,and i-STAT 1 pointof-care analyzers. Am J Clin Patho1.2008.130( 1 ) :88-92.
  • 8杨进刚.心房颤动的诊断与药物治疗(中国专家共识)[J].心脑血管病防治,2008,8(4):215-222. 被引量:119
  • 9吴晶,江蕾.临床护理路径在心房房颤华法林抗凝中的应用[J].中华现代护理杂志,2010,16(16):1892-1893. 被引量:6
  • 10邹珺,吴华章,王书平,于然,韩春雷,柯青林,刘洁,范越.凝血PT/INR快速检测服务的卫生经济学评估[J].卫生软科学,2011,25(6):367-371. 被引量:2

二级参考文献18

  • 1杨军,刘海燕,张真路,张李涛.CoaguChek XS国标标准化比值检测仪的临床应用[J].岭南心血管病杂志,2009,15(5):372-375. 被引量:4
  • 2心房颤动抗栓研究协作组,胡大一,孙艺红,张鹤萍,姜立清.华法林对非瓣膜病心房颤动抗栓的安全性和有效性研究[J].中华内科杂志,2006,45(10):800-803. 被引量:64
  • 3Denise Bockwoldt. Antithrombosis Management in Community-Dwelling Elderly: Improving Safety [J]. Geriatr Nars, 2010, (31):28-36.
  • 4John R. Petersen, Hans M. Vonmarensdorf, Heidi L. Weiss, M Tarek Elghetany. Use of error grid analysis to evaluate acceptability of a point of care prothrombin time meter[J]. Clinica Chimica Acta, 2010, (411) :131-134.
  • 5Ming Chai Kong, Teong Guan Lim, Heng Joo Ng, Yiong Huak Chan, Lai Heng Lee. Feasibility, cost-effectiveness and patients' acceptance of point-of-care INR testing in a hospital-based anticoagulation clicic[J]. Ann Hematol, 2008, (87) :905-910.
  • 6Mattias Wieloch, Andreas Hillarp, Karin Strandberg, Camilla Nilsson, Peter J. Svensson. Comparison and evaluation of a Point-of-care device(CoaguChek XS)to Owren-type prothrombin time assay for monitoring of oral anticoagulant therapy with warfarin[J].Thrombosis Research, 2009, (124):344-348.
  • 7Karen Woods, James D. Douketis, Terri Schnurr, Krystyna Kinnon, Peter Powers, Mark A. Crowther. Patient preferences for capillary vs. venous INR determination in an anticoagulation clinic: a randomized controlled trial [J]. Thrombosis Research, 2004, (114) : 161-165.
  • 8Jennifer Elston Lafata, Susan A, Martin. The cost-effectiveness of different management strategies for patients on chronic warfarin therapy[J]. Journal of General Internal Medicine, 2000, (15) :31-37.
  • 9Hankey GJ. Non-valvular atrial fibrillation and stroke prevention[J]. Med J Aust, 2001, 174(5) : 234-239.
  • 10Magdalena Sobieraj-Teague, Dolly Daniel, Barbara Farrelly, Douglas Coghlan, Alexander Gallus. Accuracy and clinical usefulness of the CoaguChek Sand XS Point of Care devices when starting warfarin in a hospital outreach setting[J].Thrombosis Research, 2009, (123): 909-913.