摘要
目的设计一种基于容积脉搏波的无袖套连续血压测量系统。方法从单一容积脉搏波中提取脉搏波传导时间,经逐步回归分析与血压建立血压估算方程,实现无创连续血压测量。结果通过对不同人群血压检测,并与鱼跃牌水银血压计进行对比,结果表明该方法和传统方法具有较好的测试一致性,测量误差优于美国医疗仪器促进协会(AAMI)推荐标准。结论该方法同传统血压测量方法相比,测量方便,可彻底摆脱缚带,并能实现无创连续测量,具有更广阔的应用前景。
Objective To develop a non-invasive continuous blood pressure measurement system without the cuff based on plethysmographic pulse wave. Methods A blood pressure estimation equation was established by the stepwise regression analysis on blood pressure and pulse wave transit time which was extracted from a single circle of plethysmographie pulse wave, and then the non-invasive continuous blood pressure measurement was realized. Results Compared blood pressure value with detection by the system and Yu-Yue brand mercury sphygmomanometer from various populations, the results indicated that the two methods exhibit good coherence, and the measurement error is better than the Association for the Advancement of Medical Instrumentation (AAMI) recommendation standard. Conclusion Compared with traditional blood pressure measurement method, the non-invasive continuous blood pressure measurement method is more convenient. It can measure blood pressure continuously without cuff and invasion, and may have promising application in the future.
出处
《航天医学与医学工程》
CAS
CSCD
北大核心
2013年第1期47-50,共4页
Space Medicine & Medical Engineering
基金
国家863重大项目(2011AA040406)
国家自然科学基金项目(61271119)
广西自然科学基金项目(2011jjA40078)
关键词
容积脉搏波
无创
连续血压测量
线性回归
plethysmographic pulse wave
non-invasive
continuous blood pressure measurement
linear regression