不同预冲方法对降低维持性血液透析患者体外循环管路中气泡和微粒污染研究 被引量:39

Study on the removal of gas and particles in extracorporeal circulation tubes for hemodialysis by physiological saline with different flow velocity
暂未订购
导出
摘要 目的评价应用不同流速、剂量的生理盐水对血液净化患者体外循环管路和血液透析器的冲洗效果。方法采用500ml、800ml生理盐水,分别以100ml/min、300ml/min和100~300ml/min的速度冲洗管路和血液透析器,检测3种方法对管路和血液透析器中气体和微粒的清除效果。结果 100ml/min冲洗速度对清除血液透析器和管路中的气体效果好,300ml/min的冲洗速度对微粒清除效果好,100~300ml/min的序贯式冲洗方法则能够同时清除血液透析器和管路中的气体和微粒。结论 100~300ml/min的序贯式冲洗方法是适用于血液透析器和管路冲洗的有效方法。 Objective To evaluate the removal efficiency of gas and particles in extracorporeal circulation tubes and filters for hemodialysis by physiological saline with different flow velocity. Methods To irrigate the tubes at the speed of 100 ml/min,300 ml/min and 100~300 ml/min with 800 ml physiological saline,and to assess the removal efficiency of gas and particles in the extracorporeal circulation tubes and filters. Results With the irrigation speed of 100 ml/min,we got the best removal result for gas; with the irrigation speed of 300 ml/min,we got the best removal result for particles; and with the sequential irrigation speed of 100~300 ml/min,we got the ideal removal result for both gas and particles. Conclusion The sequential irrigation method at 100~300 ml/min speed is the most effective way to wash the extracorporeal circulation tubes and filters for hemodialysis.
出处 《中国血液净化》 2010年第12期680-681,共2页 Chinese Journal of Blood Purification
关键词 血液透析 体外循环管路 冲洗 不溶性微粒 Hemodialysis Physiological saline Insoluble particulate