期刊文献+

植入角度及深度对种植体支抗稳定性影响的三维有限元研究 被引量:20

3D-FE study of implant anchorage stability under effects of inserting angulation and depth
暂未订购
导出
摘要 目的分析不同植入角度和植入深度下支抗种植体受力后的位移变化。方法利用pro/E软件自适应功能建立包含不同植入角度和植入深度支抗种植体和上颌骨块的三维有限元模型,在种植体顶部中央点模拟施加水平方向(与颌骨面平行,近远中方向)的正畸作用力,分析该点位移与力值变化受植入角度和植入深度影响的规律。结果①当植入深度是11mm,以30°、60°、90°植入时,需要力值分别是1 041.07、735.71、623.11 N;②当植入深度是10 mm,以30°、60°、90°植入时,需要力值分别是486.72、518.66、456.98 N;③当植入深度是9 mm,以30°、60°、90°植入时,需要力值分别是306.12、384.49、362.61N;④在本研究设计的9种工况作用下,以30°、11 mm深度植入支抗种植体时,种植体最稳定。结论深度对种植体稳定性的影响远大于角度对种植体的影响;60°倾斜度相对于90°和30°来说更加稳定。 Objective To analyze the effects of the inserting angulation and depth on the implant anchiorage stability.Methods By using Pro/E wildfire 2.0,a 3D-FEM was constructed.We gave a level force to the top point of the implant,and analyzed the relationship between its distance and the force under different inserting angulations and depths.Results The force to implants′1mm movement of different design was different:①in the depth of 11mm,with different inserting angles 30°,60°,90°,the force in sequence was:1041.07,735.71,623.11N;②in the depth of 10mm,with different inserting angles 30°,60°,90°,the force in sequence was: 486.72,518.66,456.98N;③in the depth of 9mm,with different inserting angles 30°,60°,90°,the force in sequence was: 306.12,384.49,362.61N;④It was most stable when the implant was inserted with 11mm depth and 30° angle.
出处 《口腔医学》 CAS 2010年第8期488-490,共3页 Stomatology
关键词 种植体支抗 深度 角度 三维有限元 稳定性 implant anchorage depth angle three-dimensional finite element model stability
  • 相关文献

参考文献15

  • 1Schlegel KA, Kinner F, Schlegel KD. The anatomic basis for palatal imolants inorthodontics [ J ]. Int J Adult Ortbodon Orthognath Surg,2002,17 (2) : 133 - 139.
  • 2Wilmes B,Su YY, Drescher D. Insertion angle impact on primary stability of orthodontic mini-implants [ J ].Angle Orthod, 2008,78 (6) :1065 - 1070.
  • 3孙滢滢,孔亮,刘宝林,胡开进,李德华,宋俐.圆柱种植体直径和长度的双因素分析[J].华西口腔医学杂志,2007,25(4):358-361. 被引量:6
  • 4王震东,王林,倪晓宇,马俊青.不同长度微种植体支抗应力差异的三维有限元研究[J].口腔医学,2005,25(2):96-97. 被引量:46
  • 5Miyawaki S, Koyama I, Inoue M,et al. Factors associated with the stability of titanium screws placed in the posterior region for orthodontic anchorage[ J]. Am J Orthod Dentofacial Orthod,2003,124 (4) :373 -378.
  • 6Wehrbein H, Gollner P, Diedrich P. Orthodontic load on short maxillary implants with reduced sink depth: an experimental study [ J]. Clin Oral Implants Res,2008,19(10) : 1063 - 1068.
  • 7Wilmes B, Rademacher C, Olthoff G, et al. Parameters affecting primary stability of orthodontic mini-implants [ J ]. J Orofac Orthop, 2006,67 ( 3 ) : 162 - 174.
  • 8Moon CH,Lee DG,Lee HS,et al. Factors associated with the success rate of orthodontic miniscrews placed in the upper and lower posterior buccal region [ J ]. Angle Orthod, 2008,78 ( 1 ) : 101 - 106.
  • 9姜茂庆,李德华,孔亮,邱志惠.带凹槽螺纹种植体三维有限元模型的建立[J].口腔医学研究,2007,23(2):159-161. 被引量:2
  • 10FanJianping,TangChak-Yin.THE INFLUENCE OF SHRINKAGE AND MOISTURE DIFFUSION ON IDEALIZED TOOTH STRUCTURE INVOLVING DEBONDING DAMAGE[J].Acta Mechanica Solida Sinica,2005,18(2):110-122. 被引量:2

二级参考文献25

  • 1赖红昌,熊焕国,薛淼,杨宠莹,张保卫.不同长度种植体对骨界面应力影响的三维有限元分析[J].上海口腔医学,1997,6(1):20-22. 被引量:18
  • 2Park HS.A new protocol of the sliding mechanics with micro-implant anchorage (M.I.A)[J].Korea J Orthod,2000,30 (6) : 677-685.
  • 3Park HS.An anatomical study using CT images for the implantation of micro-implants[J].Korea J Orthod,2002,32(6): 435-441.
  • 4Ivanoff C J,Grondahl K,Sennerby L,et al.Influence of variations in implant diameters:a 3-to 5-year retrospective clinical report[J].Int J Oral Maxillofac Implants,1999,14:173-180
  • 5Van Oosterwyck H,Duyck J,Vander Sloten J,et al.The influence of bone mechanical properties and implant fixation upon bone loading around oral implants[J].Clin Oral Implants Res,1998,9::407-418
  • 6del Valle V,Faulkner G,Wolfaardt J.Craniofacial osseointegrated implant-induced strain distribution:A numerical study[J].Int J Oral Maxillofac Implants,1997,12(3):200-210
  • 7Kitamura E,Stegaroiu R,Nomura S,et al.Biomechanical aspects ofmarginal bone resorption around osseointegrated implants:considerations based on a three-dimensional finite element analysis[J].Clin Oral Implants Res,2004,15 (4):401-412
  • 8Hedia HS.Stress and strain distribution behavior in the bone due to the effect of cancelous bone.Dental imphmt material and the bone height[J].Biomedical Material Eng,2002,12(2):111-119
  • 9Chun HJ,Chenog SY,Han JH,et al.Evaluation of design parameters of osseointegrated dental implants using finite element analysis[J].J Oral Rehabil,2002,29:565-574
  • 10Cehreli M,Duyck J,De Cooman M,et al.Implant design and interface force transfer.A photoelastic and strain-gauge analysis[J].Clin Oral Implants Res,2004,15(2):249-257.