摘要 基于抗坏血酸在碱性条件下与高碘酸钾和鲁米诺反应产生很强的化学发光,建立了流动注射化学发光测定抗坏血酸的新方法。在碱性条件下的氧化过程易产生过氧化物和超氧自由基,是产生化学发光的关键。该法校准曲线线性范围为6.0×10-6~4.0×10-2g/L,检出限达8×10-7g/L,相对标准偏差小于4%(n=11),采样频率为120次/h。具有灵敏度高,线性范围宽,选择性好的优点,对药用维生素C中抗坏血酸含量进行测定,得到满意结果。 The chemiluminescence reaction of ascorbic acid with potassium periodate and luminol in a alkaline solution has been studied.A novel method with flow injection for the determination of ascorbic acid was proposed , and the mechanism of reaction was also discussed .The linear range of the calibration curve is 6.0×10 -6 ~4.0×10 -2 g/L,and the detection limit is 8×10 -7 g/L with a throughput of 120 samples per hour.The RSD of determination is less than 4%( n =11).The method has been applied to the determination of ascorbic acid in pharmaceutical samples with high sensitivity and selectivity ,and satisfactory results were obtained.
出处 《分析试验室》 CAS CSCD 1999年第2期58-61,共4页 Chinese Journal of Analysis Laboratory
关键词 化学发光 流动注射 抗坏血酸 测定 Chemiluminescence flow injection ascorbic acid
  • 相关文献