期刊文献+

硫化物微生物传感器的研究 被引量:23

Study on the Microbial Sensor for Sulfide
在线阅读 下载PDF
导出
摘要 从硫铁矿的酸性土壤中分离、筛选出氧化硫硫杆菌,将其固定化制备微生物膜,再与氧电极组装成微生物传感器,用于样品中微量硫化物的测定。实验研究表明:该传感器响应S2-浓度线性范围为0.06~1.50mg/L;响应时间为3~6min;30d内测定500余次,灵敏度保持不变;取煤气站脱硫塔入口和出口煤气样品进行测定,与亚甲蓝比色法测定结果一致;取合成水样品进行回收率实验,结果为93.8%~105.0% Thiobacillus thioxidans was separated from acid soil.The microbial membrane,prepared by immobilized microorganisms,was attached to the surface of oxygen electrode to form microbial sensor for the determination of sulfide.The liner response range of the sensor is 0.06~1.50 mg/L and the response time is 3~6 min for sulfide.The sensitivity of the sensor was constant in 30 days and more than 500 assays.The microbial sensor was applied to the determination of gas samples.A good agreement was obtained between the microbial sensor and the methylene blue method.The recovery rates for sulfide added into water were 93.8%~105.0%.
出处 《河北科技大学学报》 CAS 1999年第1期10-13,共4页 Journal of Hebei University of Science and Technology
基金 河北省科委基金
关键词 微生物传感器 微生物电极 硫化物 microbial sensor microbial electrode determination of sulfide
  • 相关文献

参考文献3

  • 1国家环保局《水和废水监测分析方法》编委会.水和废水监测分析方法[M].北京:中国环境科学出版社,1989.574.
  • 2王素芳.离子选择性电极在环境监测分析中的应用[M].北京:中国环境科学出版社,1988.61-83.
  • 3孙裕生,刘宪梅,祁建珍,崔建升.微生物膜传感器BOD测定仪的研制[J].分析仪器,1992(1):1-4. 被引量:9

二级参考文献1

  • 1K. Riedel,R. Renneberg,M. Kühn,F. Scheller. A fast estimation of biochemical oxygen demand using microbial sensors[J] 1988,Applied Microbiology and Biotechnology(3):316~318

共引文献103

同被引文献510

引证文献23

二级引证文献193