种植体螺纹形状对骨界面应力分布的影响 被引量:7

Effects of different shape and angle of implant screw thread on Stress distribution in implant-bone interface
暂未订购
导出
摘要 目的研究种植体表面不同螺纹形状对种植体-骨界面应力分布的影响,以供临床筛选合适的种植系统。方法采用三维有限元法,分别对种植体施加30牛顿垂直和斜向45°两种方向的集中载荷,对不同螺纹顶角螺纹形状分别为对称、上平下斜、上斜下平的种植体-骨界面进行应力分析。结果螺纹形状为上平下斜式种植体的应力峰值较小;各组模型的最大位移值相近,但螺纹形状对称、顶角为30°者位移极值最小,为3.84×10-4mm和85.61×10-4mm。结论螺纹形状对种植体-骨界面的应力分布有影响,设计和选择种植系统时应全面考虑。 Objective To investigate the effects of different shape of implant screw thread on stress distribution in implant- bone interface. Methods Implants with different screw thread angles and shapes were divided into three groups. Group A consists of three implants with the screw thread with symmetrical shape and 30, 60, 90 degree angle, respectively. Group B consists of three implants with the screw thread with piano - slope shape and 30, 45, 60 degree angle, respectively. Group C consists of three implants with the screw thread with sloping - plane shape and 30, 45, 60 degree angle, respectively. The implants were all loaded with 30 Newton axially and 45 degree obliquely and the stress and strain in implant - bone interface were analyzed with three - dimension finite element analysis method. Results The maximal interfacial stresses for group B implants were lower than that for group A and C implants. There was no notable difference in displacement among the groups. However, displacement for group A implant with screw thread with symmetrical shape and 30 degree angle was the lowest with the value of 3.84×10^-4mm and 85.61×10^-4mm when loaded axially and obliquely, respectively. Conclusion The shape and angle of screw thread may influence the stress distribution in implant- bone interface.
出处 《现代口腔医学杂志》 CAS CSCD 北大核心 2007年第2期162-165,共4页 Journal of Modern Stomatology
关键词 种植体 螺纹 三维有限元 应力和位移 Implant Screw thread Three- dimension finite element analysis Stress and displacement
  • 相关文献

参考文献8

  • 1Esposito M,Hirsch JM,Lekholm U,et al.Biological factors contributing to failures of osseointegrated oral implants (Ⅱ).Etiopathogenesis.Eur J Oral Sci,1998,106(3):721-764.
  • 2Duyck J,Naert I.Failure of oral implants:aetiology,symptoms and influencing factors.Clin Oral Investig,1998,2 (3):102 -114.
  • 3Esposito M,Hirsch JM,Lekholm U,et al.Differential diagnosis and treatment strategies for biologic complications and failing oral implants:a review of the literature.Int J Oral Maxillofac Implants,1999,14(4):473-490.
  • 4Paul P,Grundling Nl,Joose Ch.Three-dimensional finite element model of a human mandible incorporating six osseointagrated implants for stress analysis of mandibular cantilever prosthese.Int J Oral Maxillofac Implants,1995,10:51.
  • 5Rossen IP,et al.Stress-absorbing elements in dental implants.J Prosthet Dent,1990,64:198.
  • 6王野平,单莉英,顾云峰.口腔种植技术中的工程问题探讨[J].生物医学工程学杂志,2000,17(3):328-332. 被引量:10
  • 7黄辉,马轩祥,张少锋.螺纹顶角角度对牙柱状螺旋根骨内种植体应力分布的影响[J].实用口腔医学杂志,1996,12(2):119-121. 被引量:25
  • 8兰泽栋,林珠,龙红月,周磊.支抗种植体不同螺纹顶角对骨界面应力分布的影响[J].广东牙病防治,2003,11(4):259-261. 被引量:14

二级参考文献21

共引文献40

同被引文献112

引证文献7

二级引证文献11

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部