摘要
根据串联釜模型给出样品分散度与通过管长的关系式,应用此式可以满意地解释分散度随管长的变化。
On the basis of the tanks-in-series model, an expression has been presented for the relationship between sample dispersion and reaction tubing length in the flow injection analysis. It can explain satisfactorily the change of sample dispersion with tubing length.
出处
《分析测试学报》
CAS
CSCD
1996年第3期82-84,共3页
Journal of Instrumental Analysis
关键词
流动注射分析
分散度
管长
串联釜模型
Flow injection analysis, Dispersion. To whom correspondeoce should be addressed.