氟化锂(镁铜磷)元件的剂量学特性研究 被引量:2

Study on thermoluminescence characteristics of LiF∶Mg,Cu,P
暂未订购
导出
摘要 目的 考查GR 2 0 0A型LiF :Mg,Cu ,P(氟化锂∶镁 ,铜 ,磷 )元件在常规剂量监测中的适用性。方法 使用RGD 3A型热释光测读仪对元件的有关剂量学特性进行检测。结果 元件分散性误差为 4%;2mGy~ 17mGy范围内 ,元件的剂量响应具有很好的线性 ;元件的重复性偏差为 2 %;32 0℃退火时 ,元件灵敏度最低 ;测量结果基本不受加热速率的影响。结论 该LiF∶Mg ,Cu ,P热释光元件完全符合个人剂量常规监测要求。 Object To evaluate the adaptability of GR-200A chips which will be used to monitor personnel dose.Methods The tests were carried out to measure relative characteristics of these elements by using RGD-3A dosimetric reader.Results The result shows that the distribution of this kind of dosimeters is 4%, its linearity of dose response is good, and the deviation of result is 2% after repeated irradiation. In addition, the sensitivity of LiF∶Mg,Cu,P is reaches the bottom at 320℃, but it is hardly affected by the heating rate.Conclusion Above results indicated that this kind of TLD has excellent dosimetric characteristics which allow its introduction in routine dosimetry application.
出处 《中国卫生工程学》 CAS 2003年第1期4-6,共3页 Chinese Journal of Public Health Engineering
关键词 氟化锂 热释光 LiF:Mg Cu P元件 剂量学特性 thermoluminescence LiF∶Mg,Cu,P dosimetric characteristics
  • 相关文献

参考文献6

  • 1[1]G Cai,K Geng,Q Wang.The environmental monitoring of the natural radiation background in Antarctic with LiF:Mg,Cu,P TLD and X-γ radiometry.Radiation Protection Dosimetry,1995,60(3):259 ~ 262.
  • 2[2]Tso Wong Man-yin.Thermoluminescence authenticity assay of Chinese pottery in Hong Kong,Nuclear Techniques,1989,12 (10):594 ~ 595.
  • 3[3]M.Mosonvitch.Personnel dosimetry using LiF:Mg,Cu,P.Radiation Protection Dosimetry,1999,85(1 ~ 4):49 ~ 56.
  • 4[4]L.Aurrivan,D.Spain,H.Donndllyard,et al.A thermoluminescence doaimetry system for personal monitoring in Ireland.Radiation Protection Dosimetry,2001,1996(1 ~ 3):209 ~ 212.
  • 5[5]Kaiyong Tang,Honging Zhu,Wenxiu Shen.Thermal loss and recovery of thermoluminescence sensitivity in LiF:Mg,Cu,P.Radiation Protection Dosimetry,2000,90(4):449 ~ 452.
  • 6[6]A.S.Praohan.Influonce of heating rate on the TL response of LiF TLD-700,LiF:Mg,Cu,P and Al2O3:C.Radiation Protection Dosimetry,1995,158(3):205~209.

同被引文献11

引证文献2

二级引证文献18

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部