摘要
高压介质损耗因数标准器是实现介质损耗测试仪校准的基础,通常采用电容电阻组合实现,存在量值范围受限、稳定性不足、成本较高等问题。本文基于正交幅值调制的信号移相法,研制了一种标准器样机,并采用电桥比较法进行实验验证。结果表明:该标准器在(15~50)Hz范围内,可实现电容量(100~500)nF和损耗因数(0.00000~0.1)的任意量值输出,电容量允许最大误差为±0.2%,损耗因数的允许最大误差为±(0.5%×读数+0.00005)。
The high-voltage dielectric loss factor standard is the basis for calibrating dielectric loss testers,usually using a combination of capacitors and resistors,which has problems such as limited range of values,insufficient stability,and high cost.This article is based on the signal phase shifting method of orthogonal amplitude modulation,and develops a prototype of a standard instrument,which is experimentally verified using the bridge comparison method.The results show that the standard can achieve arbitrary output values of capacitance(100~500)nF and loss factor(0.00000~0.1)within the range of(15~50)Hz.The maximum allowable error for capacitance is±0.2%,and the maximum allowable error for loss factor is±(0.5%X reading+0.00005).
作者
王一帆
孙明
王异凡
龚金龙
赵琳
郑迪
WANG Yifan;SUN Ming;WANG Yifan;GONG Jinlong;ZHAO Lin;ZHENG Di
出处
《计量与测试技术》
2025年第5期8-11,共4页
Metrology & Measurement Technique
基金
国网浙江省电力有限公司科技项目(项目名称:基于多信息融合感知与量值溯源的低频输变电设备状态评估关键技术研究,项目编号:5211DS23000B)。
关键词
介质损耗因数
正交幅值调制
信号移相
标准器
不确定度分析
dielectric loss factor
quadrature amplitude modulation
signal phase shift
standard device
uncertainty analysis