期刊文献+

ATP生物荧光法在达芬奇机械臂清洗效果检测中的应用 被引量:20

暂未订购
导出
摘要 目的观察ATP生物荧光法在达芬奇机械臂清洗效果检测中的应用效果。方法采用便利抽样方法抽取泌尿外科前列腺手术后机器人机械臂15套,每套次再用便利抽样方法抽取机械臂4把(电凝剪、电凝钩、针持、无损伤钳各1把),手工加超声波清洗后,采用10倍放大镜目测和ATP生物荧光法两种方法检测清洗效果,当两种检测结果不一致时采用残留蛋白检测法检测。结果60件机械臂放大镜目测法检查清洗合格率为95.00%,ATP生物荧光法检测清洗合格率为78.33%,两种检测法合格率比较差异有统计学意义(P<0.05),对两种检测结果不一致的10件机械臂经残留蛋白清洗测试棒检测均为不合格。结论ATP生物荧光法检测达芬奇机械臂清洗效果较放大镜目测法好。
出处 《护理与康复》 2020年第9期71-73,共3页 Journal of Nursing and Rehabilitation
基金 浙江省医药卫生科技计划项目,编号2018ZH010。
  • 相关文献

参考文献3

二级参考文献43

  • 1王华生,孙雪莹,梁树森,徐然.影响医疗器械清洗质量的因素[J].中华医院感染学杂志,2007,17(5):553-554. 被引量:117
  • 2Davidson CA, Griffith C J, Peters AC, et al. Evaluation of two methods for monitoring surface cleanliness - ATP bioluminescence and traditional hygiene swabbing [J]. Luminescence, 1999, 14 (1) :33 -38.
  • 3Alfa M J, Fatima I, Olson N. The adenosine triphosphate test is a rapid and reliable audit tool to assess manual cleaning adequacy of flexible endoscope channels [ J ]. Am J Infect Control,2013, 41 (3) :249 -253.
  • 4Hansen D, Benner D, Hilgenhoner M, et al. ATP measurement as method to monitor the quality of reprocessing flexible endo- scopes [J]. Ger Med Sci J, 2004,(2) :04.
  • 5Obee PC, Griffith C J, Cooper RA, et al. Real - time monitoring in managing the decontamination of flexible gastrointestinal endo- scopes [ J ]. Am J Infect Control,2005,33 (4) :202 - 206.
  • 6Fushimi R, Takashina M, Yoshikawa H, et al. Comparison of a- denosine triphosphate, microbiological load, and residual protein as indicators for assessing the cleanliness of flexible gastrointesti- nal endoscopes [ J]. Am J Infect Control, 2013,41 ( 2 ) : 161 - 164.
  • 7Alfa MJ, Olson N, Murray BL. Comparison of clinically relevant benchmarks and channel sampling methods used to assess manual cleaning compliance for flexible gastrointestinal endoscopes [ J]. Am J Infect Control ,2014,42 ( 1 ) : 1 - 5.
  • 8Marena C, Lodola L, Zecca M, et al. Assessment of handwash- ing practices with chemical and microbiologic methods: prelimi- nary results from a prospective crossover study [ J ]. Am J Infect Control, 2002,30 (6) :334 - 340.
  • 9Willis C1, Morley R, Westbury J,et al. Evaluation of ATP biolu- minescence swabbing as a monitoring and training tool for effec- tive hospital cleaning [J]. Brit J Infect Control,2007,8 (5) :17 -21.
  • 10Boyce JM, Havill NL, Dumigan DG, et al. Monitoring the effec- tiveness of hospital cleaning practices by use of an adenosine triphosphate bioluminescence assay[ J]. Infect Cont Hosp Epide- miol, 2009,30(7) :678 -684.

共引文献37