一种用于平板型FAIMS的电晕放电离化源 被引量:3

A Corona Discharge Ionization Source for the Planar FAIMS
在线阅读 下载PDF
导出
摘要 本文制备了一种适用于平板型FAIMS的电晕放电离化源,由电晕针、钼网、铜底座和聚四氟乙烯外筒等部件组成。其中电晕针使用直径为0.5 mm的钨丝通过电解加工方法制备,加工条件为:电解电压6 V;电解液为NaOH,1 mol/L;制备时间控制在160 s。观测了离化源在针-网距离为2 mm^8 mm时的伏安特性,针-网距离越远放电起始电压越高,放电电压(kV)与电流/电压(μA/kV)呈线性关系。测量了1μA^40μA放电电流下,FAIMS检测的甲基膦酸二甲酯DMMP(Dimethyl methylphosphonate)样品的输出响应图谱,反应离子峰与DMMP离子峰都随放电电流增加而增大。 A corona discharge ionization source is designed for planar FAIMS,which consists of corona needle,molybdenum mesh,copper base and PTFE cylinder. The corona needle is made from tungsten wire with a diameter of0.5 mm through electrolytic processing with conditions setting as electrolysis voltage at 6 V,the electrolyte Na OH( 1 mol/L) and duration time160 s. The voltage-current characteristics of the corona discharge ionization source were observed at pin-grid distance between 2 mm ~ 8 mm,where the correlation between discharge voltage( kV) and the current/voltage( μA/kV) is approximately linear. The FAIMS spectra of DMMP were measured at discharge currents between 1 μA ~ 40 μA,which can be observed that both the reaction ion peak and the DMMP ion peak increased with the discharge current.
作者 赵东杰 刘军 丁庆行 周明 马廷伟 ZHAO Dongjie;LIU Jun;DING Qingxing;ZHOU Ming;MA Tingwei(School of Information,Beijing Wuzi University,Beijing 101149,China;Beijing Higher Institution Engineering Research Center of Logistics Technology,Beijing Wuzi University,Beijing 101149,China)
出处 《传感技术学报》 CAS CSCD 北大核心 2018年第6期887-890,共4页 Chinese Journal of Sensors and Actuators
基金 北京高校物流技术工程研究中心项目(BJLE2010) 北京物资学院青年科研基金项目(2014XJQN15)
关键词 高场非对称波形离子迁移率谱 电晕放电 离子源 甲基膦酸二甲酯 FAIMS Corona Discharge Ionization source DMMP
  • 相关文献

参考文献4

二级参考文献33

  • 1邵士勇,阚瑞峰,侯可勇,李海洋.离子迁移谱仪的研究进展[J].现代科学仪器,2004,21(4):9-12. 被引量:20
  • 2郭希山,童基均,陈裕泉,杨祥龙.用于室内有毒气体快速检测的便携式CC/SAW电子鼻[J].传感技术学报,2006,19(1):68-73. 被引量:8
  • 3Kolakowski B M, Mester Z. Review of Applications of High-Field Asymmetric Waveform Ion Mobility Spectrometry ( FAIMS ) and Differential Mobility Spectrometry(DMS) [ J]. Analyst,2007,132 : 842-864.
  • 4Shvartsburg A A, Tang K Q, Smith R D. Differential Ion Mobility Separations of Peptides with Resolving Power Exceeding 50 [ J ]. Analytical Chemistry ,2010,82:32-35.
  • 5Shvartsburg A A, Smith R D. Optimum Waveforms for Differential Ion Mobility Spectrometry (FAIMS) [ J ]. Journal of the American Society for Mass Spectrometry,2008,19 : 1256-1295.
  • 6Papanastasiou D, Wollnik H, Rico G, et al. Differential Mobility Separation of Ions Using a Rectangular Asymmetric Waveform[ J ]. Journal of Physical Chemistry A ,2008,112:3638-3645.
  • 7Krylov E V, Coy S L, Vandermey J, et al. Selection and Generation of Waveforms for Differential Mobility Spectrometry [ J ]. Rev Sci Instrum,2010,81:24101.
  • 8Krylov E V,Pulses of Special Shapes Formed on a Capacitive Load [ J ]. Instruments and Experimental Techniques, 1997 ,40 :628-631.
  • 9Arce L, Menendez M, Garrido-Delgado R, et al. Sample-Introduction Systems Coupled to Ion-Mobility Spectrometry Equipment for Determining Compounds Present in Gaseous, Liquid and Solid Samples[J].Trac-Trends in Analytical Chemistry ,2008 ,27 :139-150.
  • 10Krylov E V, Nazarov E G, Miller R A. Differential Mobility Spectrometer : Model of Operation [ J ]. International Journal of Mass Spectrometry, 2007,266 : 76- 85.

共引文献12

同被引文献17

引证文献3

二级引证文献7

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部