锥形束CT图像引导下宫颈癌患者放疗摆位误差的临床应用分析 被引量:9

Analysis of Setup Errors on Cone-Beam CT-Based Image Guided Radiation Therapy for Cervical Cancer
暂未订购
导出
摘要 目的分析千伏级锥形束CT引导下以真空垫为体位固定装置的宫颈癌患者图像引导放疗中的摆位误差。方法 50例宫颈癌患者分别于放疗第1周、第3周、第5周采用瓦里安直线加速器机载影像系统行锥形束CT(CBCT)扫描,与计划CT图像匹配获得左右、头足、腹背及旋转方向的摆位误差,并计算计划靶体积(PTV)外放边界。结果 50例患者共行150次扫描,左右、头足、腹背及旋转方向的摆位误差分别为(2.89±1.63)mm、(4.68±2.19)mm、(2.24±1.36)mm、(7.62±5.77)deg,左右、头足、腹背方向PTV外放边界分别为8.36、13.23、6.55 mm。结论应用锥形束CT扫描可测量宫颈癌患者的摆位误差并在线校正,以减少误差;还可估算PTV外扩边界值,为临床医师提供治疗依据。 Objective To evaluate setup errors of image guided radiation therapy( IGRT) for cervical cancer with vacuum cushions using kilovolt cone beam CT( KVCBCT). Methods Varian-IX linear accelerator on-board imaging system was used to acquire CBCT scans before irradiation of 50 cervical cancer patients in the 1st,3rd,and 5th week,respectively.CBCT scans and planning CT scans were auto-matched to obtain the setup errors of left-right( x),superior-inferior( y),anterior-posterior( z) and rotation( r). The value of the MPTV was calculated,meanwhile. Results 50 patients acquired 150 CBCT scans. The setup errors in x-,y-,z-,r-dimension were( 2. 89 ± 1. 63) mm、( 4. 68 ± 2. 19) mm、( 2. 24 ± 1. 36) mm、( 7. 62 ±5. 77) deg. The MPTV were 8. 36、13. 23、6. 55 mm in x-,y-,z-dimension. Conclusions CBCT on line correction could reduce the setup errors of cervical cancer patients,and estimate the MPTV,thereby provide the basis for clinicians.
出处 《中南医学科学杂志》 CAS 2015年第2期201-203,共3页 Medical Science Journal of Central South China
关键词 体层摄影术 锥形束 图像引导放疗 摆位误差 宫颈癌 Tomography cone-beam Image guided radiotherapy Setup errors Cervical cancer
  • 相关文献

参考文献11

  • 1戴建荣,胡逸民.图像引导放疗的实现方式[J].中华放射肿瘤学杂志,2006,15(2):132-135. 被引量:184
  • 2Stroom JC,de Boer HC,Huizenga H,et al. Inclusion of geo- metrical uncertainties in radiotherapy treatment planning by means of coverage probability [ J ]. Int J Radiat Oncol Biol Phys, 1999,43 ( 4 ) : 905-919.
  • 3Stroom JC, Heijmen BJ. Geometrical uncertainties, radio- therapy planning margins, and the ICRU-62 report [ J ]. Ra- diother Oncol,2002,64 ( 1 ) :75-83.
  • 4van Herk M, Remeijer P, Rasch C, et al. The probability of correct target dosage: dose-population histograms for deri- ving treatment margins in radiotherapy [ J ]. Int J Radiat On- col Biol Phys ,2000,47 (4) : 1121-1135.
  • 5van Herk M. Errors and margins in radiotherapy[ J]. Semin Radiat Oncol, 2004,14 ( 1 ) : 52-64.
  • 6曹晓辉,刘明,翟福山,王安峰,杨永锋,刘兵,包超恩.用锥形束CT图像引导技术分析肺癌放疗摆位误差及其对剂量分布影响[J].中华放射肿瘤学杂志,2012,21(2):163-166. 被引量:19
  • 7林益匡,缪龙华,张忠金.胸腹部肿瘤适形放射治疗摆位误差临床分析[J].福建医药杂志,2012,34(1):34-36. 被引量:5
  • 8Fu W, Yang Y, Li X, et al. Dosimetric effects of patient ro- tational setup errors on prostate IMRT treatments [ J ]. Phys Med Biol,2006,51 (20) :5321-5331.
  • 9Astreinidou E, Bel A, Raaijmakers CP, et al. Adequate mar- gins for random setup uncertainties in head-and-neck IMRT [ J]. Int J Radiat Oncol Biol Phys ,2005,61 (3) :938-944.
  • 10陈国章,何晓华,林茂盛,罗志强,林承光,张黎.增加摆位参考点对腹部肿瘤真空袋固定摆位重复性价值的研究[J].肿瘤学杂志,2008,14(9):760-762. 被引量:2

二级参考文献45

共引文献224

同被引文献74

引证文献9

二级引证文献36

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部