摘要
介绍了发光粒子绝对浓度的一种简单易行的方法—体光源模拟标定法.体光源模拟标定法测量发光粒子的绝对浓度,是把发光气体以某一流速引入一已知体积的流动光池中,再通过具有低象差失真的光学系统,把该体光源成像在探测器的有效表面上.探测器和测量仪器组成的测量系统,要经过标准光源和电学标定.本方法用于单重态氮O2(1△)绝对浓度的测量,测出其绝对浓度和分压,其测量结果的相对误差小于20%.
A calibrated piston source of light, which simulates a cylindrical-volume luminosity source, has been introduced to measure the absolute concentration of illuminant particles. It is proved that this Piston Source Method is one of the simplest and most convenient ways to measure the absolute concentration of illuminant particles, especially in real time measurements of O2(a1△) in a singlet oxygen generator. The relative error of O2:(a1△) concentration is less than 20%.
出处
《量子电子学报》
CAS
CSCD
北大核心
2001年第4期309-314,共6页
Chinese Journal of Quantum Electronics
关键词
绝对浓度
体光源模拟法
发光粒子
absolute concentration
illuminant particles
piston source method