三种光学测量仪器与A超对近视眼中央角膜厚度测量的对比研究 被引量:9

Comparison of central corneal thickness measured by three optical measurement devices with ultrasound pachymetry
暂未订购
导出
摘要 目的:探讨眼前段光学相干断层扫描仪( Visante OCT )、裂隙灯型眼前段光学相干断层扫描仪( SL-OCT )、OrbscanⅡz角膜地形图/角膜测厚系统( OrbscanⅡz)和A型超声测量仪测量近视人群中央角膜厚度( central corneal thickness ,CCT)的差异及相关性,以评价三种光学测量仪器在测量中央角膜厚度的准确性,为其临床应用提供依据。 方法:对近视患者43例86眼分别用Visante OCT、SL-OCT、OrbscanⅡz和A型超声测量仪进行CCT的测量,并对四种仪器测量的CCT值采用单因素方差分析, LSD-t法两两比较,Pearson相关分析。 结果:Visante OCT、SL-OCT、OrbscanⅡz和A型超声测量仪测量 CCT 的平均值分别为(521.05±29.88)μm、(533.73±31.83)μm、(534.02±32.96)μm、(540.60±31.11)μm,95%可信区间分别为(514.57~527.54),(526.83~540.65),(533.85~547.35),(536.75~544.45)。 Visante OCT 与SL-OCT(P=0.009)、A型超声测量仪(P<0.01)、OrbscanⅡz(P=0.008)的CCT测量值差异均有统计学意义。 Visante OCT 的CCT测量值分别低于SL-OCT的CCT测量值(12.69±9.32)μm,低于A型超声测量仪的 CCT 测量值(19.55±8.64)μm,低于OrbscanⅡz的CCT测量值(12.96±12.21)μm。 SL-OCT与A型超声测量仪( P=0.158)、OrbscanⅡz ( P=0.954)的CCT测量值差异均无统计学意义。 OrbscanⅡz与A型超声测量仪的 CCT 测量值差异无统计学意义( P=0.176)。四种方法CCT测量值均呈正相关( P<0.01)。 结论:三种光学测量仪器测量CCT值与A超测量值均有很好的相关性。 OrbscanⅡz和SL-OCT测量CCT值与A超的测量值基本一致。三种光学测量仪器与A超相比具有非接触性、无创性、测量时间短等特点,更易被患者接受。 AIM:To compare the difference and correlation of central corneal thickness ( CCT ) measured by Zeiss Visante Anterior Segment Optical Coherence Tomography ( Visante OCT ) , Heidelberg Slit -Lamp Anterior Segment Optical Coherence Tomography ( SL-OCT) , OrbscanⅡz Computerized Topography ( OrbscanⅡz ) and A-mode ultrasound pachymetry. METHODS:Forty-three patients with myopia (86 eyes) were included. Sequential measurements of CCT was performed by the same operator with two different anterior segment OCT devices ( VisanteTM , Carl Zeiss, Meditec and Heidelberg EngineeringTM ), OrbscanⅡz Computerized Topography ( Bausch and Lomb IncTM ) and A-mode ultrasound pachymetry.The consistency of measurements was statistically analyzed by one-way ANOVA and LSD pair-wise comparisons.The differences were analyzed by Pearson correlation coefficients. RESULTS:The average CCT measured by Visante OCT, SL -OCT, Orbscan Ⅱ z, and A -mode ultrasound pachymetry were (521.05 ±29.88)μm, (533.73 ±31.83)μm, (534.02 ±32.96 )μm, and ( 540.60 ±31.11 )μm, the 95%confidence interval were (514.57 ~527.54), (526.83 ~540.65), (533.85~547.35), (536.75~544.45).There was statistically significant difference of CCT between using of Visante OCT and SL-OCT ( P =0.009 ). There was statistically significant difference of CCT using of Visante OCT and A-mode ultrasound pachymetry ( P〈0.01 ), OrbscanⅡ z ( P =0.008 ). The Visante OCT CCT measurements was significantly less than CCT with SL-OCT ( 12.69 ±9.32 ) μm. Compare with A -mode ultrasound pachymetry and OrbscanⅡz, the Visante OCT measurements significantly underestimated central corneal thickness by a mean of ( 19.55 ±8.64 )μm and (12.96±12.21)μm.There was no significant difference of CCT between using of SL-OCT and A-mode ultrasound pachymetry (P=0.158), OrbscanⅡz (P=0.954).There was no significant difference of CCT between using of OrbscanⅡz and A-mode ultrasound pachymetry ( P=0.176 ) . There were positive correlationship among Visante OCT, SL-OCT, OrbscanⅡz, and A-mode ultrasound pachymetry in CCT measurement ( P 〈0.01 ) . CONCLUSION: There are excellent correlationship among Visante OCT, SL-OCT,OrbscanⅡz, and A-mode ultrasound pachymetry in CCT measurement.The data shows that the CCT obtained by SL-OCT and OrbscanⅡz do not have significant difference as compared with that from the A-mode ultrasound pachymetry.Three optical measurement devices is more likely to be accepted by patients because of its non-contact, non-invasive, and short measurement time as compared with the A-mode ultrasound pachymetry.
出处 《国际眼科杂志》 CAS 2013年第12期2545-2548,共4页 International Eye Science
  • 相关文献

参考文献15

  • 1下勤美,陈跃同.屈光手术学.北京:人民卫生出版社,2004:35.
  • 2Medeiros FA, Sample PA, Weinreb RN. Corneal thickness measurements and visual function abnormalities in ocular hypertensive patients. Am J Ophthalmol 2003 , 135 ( 2 ) : 131 - 137.
  • 3Marsieh MW, Bullimore MA. The repeatability of corneal thickness measures. Cornea 2000 , 19 ( 6 ) :792-795.
  • 4Solomon OD. Corneal indentation during ultrasonicpachometry. Cornea 1999 , 18(2 ) :214-215.
  • 5Asensio I, Rahhal SM, Alonso L, et al. Corneal thickness values before and after oxybuprocaine 0.4% eye drops. Cornea 2003 , 22 ( 6 ) :527-532.
  • 6Chakrabarti HS, Craig JP, Brahma A, et al . Comparison of corneal thickness measurements using ultrasound and Orbscan slit- scanning topography in normal and post- LASIK eyes. J Cataract Refract Surg 2001 ,27( 11 ) : 1823-1828.
  • 7Gonzalez-Meijome JM, Cervino A, Yebra-Pimentel E, et al . Central and peripheral corneal thickness measurement with Orbscan Ⅱ and topographical ultrasound pachymetry. J Cataract Refract Surg 2003,29 (1) :125-132.
  • 8Izatt IA, Flee MR, Swanson EA, et al. Micrometer-scale resolution imaging of the anterior eye in vivo with optical coherence tomography. Arch Ophthalmol 1994 , 112( 12 ) : 1584-1589.
  • 9Wirbelauer C, Scholz C, Haberle H, et al. Corneal optical coherence tomography before and after phototherapeutic keratectomy for recurrent epithelial erosions(2). J Cataract Refract Surg 2002,28(9):1629-1635.
  • 10Hoerauf H, Gordes RS, Seholz C, et al. First experimental and clinical results with transscleral optical coherence tomography. Ophthalmic Surg Lasers 2000 ,31 ( 3 ) : 218-222.

二级参考文献3

共引文献1

同被引文献88

引证文献9

二级引证文献25

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部