期刊文献+

BEGe探测器晶体参数的蒙特卡罗计算与实验表征技术研究 被引量:5

Study on Parameter Optimization of Planar BEGe Detector Using Monte Carlo Simulation and Experiment
在线阅读 下载PDF
导出
摘要 利用241 Am发射的59.54keV平行γ射线束对宽能BEGe探测器晶体进行扫描测量,通过晶体端面垂直扫描确定晶体直径,通过晶体侧面垂直扫描确定晶体高度。利用厂家提供的参数以及扫描得到的晶体参数建立蒙特卡罗计算模型,再通过点源和体源的实验测量值对晶体前端面及侧面死层厚度进行表征调节,使其与实验值符合,从而获得最佳理论计算模型。结果表明,获得的参数正确可靠,其点源探测效率的理论计算结果与实验测量值在2%内符合。该方法极大改善了蒙特卡罗计算BEGe晶体探测效率的精度,可应用于探测器的效率刻度及环境放射性调查。 Based on BEGe γ spectrometer, the scanning measurement was operated by parallel 59.54 keV y-ray which was emitted by 241Am. The crystal diameter was obtained by scanning perpendicular crystal front surface, and the crystal height was obtained by scanning perpendicular crystal lateral surface. The crystal parameters provided by manufacture and scanning were used to establish the Monte Carlo module. Based on the calibration results of the sources, the front and lateral dead layer thinknesses of BEGe were characterized. A good consistent result between Monte Carlo result and experiment value was finished by using character parameters, and the best theory module was obtained. Viewing the result data, the parameters are valid and reliable. The Monte Carlo results are consistent with the experimental results of point source in 2%. Therefore, this method accuracy, and it can be widely applied in radioactive investigation.
出处 《原子能科学技术》 EI CAS CSCD 北大核心 2013年第8期1411-1416,共6页 Atomic Energy Science and Technology
关键词 扫描测量 蒙特卡罗模拟 晶体参数 scanning measurement Monte Carlo simulation crystal parameter
  • 相关文献

参考文献6

  • 1Luis R, BENTO], CARY ALHAL G, et al. Parameter optimization of a planar BEGe detector using Monte Carlo simulations[J]. Nuclear In?struments and Methods in Physics Research A, 2010, 623: 1014-1019.
  • 2PETIT G L,JUTIER C. GROSS P. et al. Low?level activity measurement of 131 Xe?", 133 Xe'" , 135 Xe and 133 Xe in atmospheric air samples using high-resolution dual X-y spectrometry[J]. Ap?plied Radiation and Isotopes, 2006, 64: 1 307- 1 312.
  • 3BOSONJ, AGREN G.JOHANSSON L, et al. A detailed investigation of HPGe detector re?sponse for improved Monte Carlo efficiency cal?culations[J]. Nuclear Instruments and Methods in Physics Research A, 2008, 587: 304-314.
  • 4SCHLAGER M. Precise modeling of coaxial ger?manium detectors in preparation for a mathemati?cal calibration[J]' Nuclear Instruments and Methods in Physics Research A, 2007,580: 137- 140.
  • 5MKEYSER R. Resolution and sensitivity as a function of energy and incident geometry for ger?manium detectors[J]. Nuclear Instruments and Methods in Physics Research S, 2004, 213: 236- 240.
  • 6田自宁,贾明雁,李惠彬,成智威,巨凌军,申茂泉,杨晓燕,晏林,冯天成.土壤样品γ谱分析自吸收校正技术研究[J].辐射防护,2010,30(1):54-62. 被引量:13

二级参考文献6

共引文献12

同被引文献10

引证文献5

二级引证文献17

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;