期刊文献+

高效液相色谱法测定杀虫环含量的研究 被引量:2

Determination of thiocyclam content by Reversed Phase HPLC
在线阅读 下载PDF
导出
摘要 【目的】建立了反相高效液相色谱法测定杀虫环含量的方法。【方法】采用反相C18烷基柱,以甲醇一磷酸水溶液(体积比40:60)为流动相,柱温为30℃,流速为1.0mL/min,采用UVW紫外检测器在260m处进行检测,外标法定量。[结果]方法峰面积与质量浓度的线性关系良好,平均回收率100.10%,标准偏差为0.15%。[结论]该方法简便、快速、分离效果好、精密度和准确度高、线性关系好。 The sample was separated on a diamocil C18 column at the temperature of 30 ℃C with the elution of methanol and phosphoric acid solution (40: 60, by vol) as mobile phase at a flow rate of 1.0 mL/min. The detection wavelength was 260 nm, quantified by external standard method. [Resuits] The assay exhibited a good linearity, the average recovery was 100.10%, and the standard deviation was 0.15%. [Conclusions] The method has the advantages of convenience, fast speed, good separation, high precision and accuracy, and good linear correlations.
作者 周志慧
出处 《中国化工贸易》 2013年第6期279-280,共2页 CHINA CHEMICAL TRADE
关键词 反相高效液相色谱 杀虫环 杀虫剂 reversed phase HPLC thiocyclam insecticide