多中心临床研究:采用X线衰减校正提高^(99m)Tc-3PRGD_2肺癌SPECT图像定量分析精确性 被引量:1

Multicenter clinical study:the use of X-ray attenuation correction to improve quantitative analysis precision for ^(99m)Tc-3PRGD_2 lung cancer SPECT image
暂未订购
导出
摘要 目的:采用X线衰减校正提高前瞻性、多中心研究中99mTc-3PRGD2肺癌SPECT图像定量分析的精确性。方法:经病理确诊的肺癌患者27例,其中肺腺癌23例、鳞癌4例。患者静脉注射99mTc-3PRGD2后40分钟顺序进行全身和局部断层扫描。采用迭代图像重建方法,对获得的99mTc-3PRGD2SPECT图像进行X线衰减校正(CTAC)和非衰减校正重建(NAC)。测量CTAC和NAC病灶所在层面的总计数,感兴趣区域(ROI)病灶的总计数、最大计数和平均计数。分别以病灶ROI内最大计数的38%、42%和48%作为阈值测量CTAC和NAC图像病灶最大长径。结果:经过CTAC后病灶所在层面总计数增加2.09倍,ROI区域病灶的总计数增加3.32倍,病灶最大计数和平均计数与本底比值分别提高了65.9%和70.3%。以病灶ROI最大计数42%作为阈值,CTAC后的病灶最大长径为(5.65±2.77)cm,与CT测量结果一致。结论:采用X线衰减校正是提高99mTc-3PRGD2定量分析的有效方法。 Objective: Using X-ray attenuation correction to improve a prospective,multicenter study of 99mTc-3PRGD2 SPECT image quantitative analysis precise in lung patients.Methods: 27 patients,adenocarcinoma of 23 cases and squamous cell carcinoma of 4 cases,diagnosed by pathology.Images of whole body scan and local tomography were obtained after 99mTc-3PRGD2 injection of 40 minutes intravenously.Iterative image reconstruction method was used to obtain the X-ray attenuation correction(CTAC) and no attenuation correction reconstruction(NAC) on the basis of 99mTc-3PRGD2 SPECT images.Total count of the lesions slice,total count,maximum and average count of the region of interest(ROI) were extracted individually from the slices of CTAC and NAC frames.The maximum count 38%,42% and 48% within ROI were chosed as the threshold to measure the diameter length of CTAC and NAC image focus.Results: The results showed that after CTAC,the total count,of the focus slice was increased 2.09 times,the total count of the ROI was increased 3.32 times,maximum and mean counts and the background ratio were increased by 65.9% and 70.3%,individually.With the maximum count 42% of the lesion ROI as the threshold,the lesions maximum size was of(5.65±2.77) cm after CTAC,which is coincidence with CT measurement results.Conclusion: Using X-ray attenuation correction helps to improve 99mTc-3PRGD2 quantitative analysis.
出处 《中国临床医学影像杂志》 CAS 2013年第5期321-324,共4页 Journal of China Clinic Medical Imaging
基金 国家自然科学基金项目(81171369)
关键词 肺肿瘤 体层摄影术 发射型计算机 单光子 Lung neoplasms Tomography emission-computed single-photon
  • 相关文献

参考文献7

  • 1Philipp R, Hans V, Joachim H, et al. Absolute quantification in SPECT[J]. Eur J Nucl Med Mol Imaging, 2011, 38(Suppl): S69-S77.
  • 2Alain S, Daniel N, Claire B. Quantitative capabilities of four state-of-the an SPECT-CT cameras[J]. EJNMMI Res, 2012, 2(1): 45-70.
  • 3孟庆乐,姚薇萱,王峰,等.X线衰减校正对SPECT断层图像质量的影响[J].中国临床医学影像杂志,2010,18(5):403-406.
  • 4闫新慧.心肌灌注SPECT显像采用X线CT衰减校正的临床应用[J].中国临床医学影像杂志,2010,21(6):388-391. 被引量:2
  • 5Erdi YE, Mawlawi O, Larson SM, et al. Segmentation of lung lesion volume by adaptive positron emission tomography image thresholding[J]. Cancer, 1997, 80(12): 2505-2509.
  • 6Lamon SM, Erdi Y, Akhurst T, et ah Tumor treatment response based on visual and quantitative changes in global tumor glycolysis using PET-FDG imaging: the visual response score and the change in total lesion glycolysis [J]. Clin Positron Imaging, 1999, 2(3): 159-171.
  • 7Remy L, Anne E, Lee Nancy Y, et al. ^18F-FDG PET/CT metabolic tumor volume and total lesion glycolysis predict outcome in oropharyngeal squamous cell careinoma [J]. J Nuel Med, 2012. 53(10): 1506-1513.

二级参考文献8

  • 1Kial H,Van Train KF,Friedman JD,et al.Quantilative stress-redistribution thallium-201 SPECT using prone imaging methodological development and validation[J].J Nucl Med,1992,33:1509-1515.
  • 2Garcia EV.SPECT attenuation correction:An essential tool to realize nuclear cardiology's manifest destiny[J].J Nucl Cardiol,2007,14:16-21.
  • 3Gaemperli O,Scbepis T,Valenta I,et al.Cardiac image fusion from stand-alone SPECT and CT clinical experience[J].J Nucl Med,2007,48:696-703.
  • 4Segall GM,Davis MJ.Prone versus supine TI myocardial SPECT:a method to decrease artifactual inferior wall defects[J].J Nucl Med,1989,30:548-555.
  • 5Jean PE,Florent JC,Serge JM,et al.Prone decubitus:a solution to inferior wall attenuation in 201TL myocardial tomography[J].J Nucl Med,1989,30:398-401.
  • 6Gerrnano G,Slomka PJ,Barnan DS.Attenuation correction in cardiac SPECT:the boy who cried wolf[J].J Nucl Cardiol,2007,14:25-35.
  • 7谭天秩.临床核医学[M].北京:人民卫生出版社,1992.63-568.
  • 8刘斌,曾宇,黄蕤,匡安仁.心肌灌注显像CT衰减校正伪影产生的初步研究[J].华西医学,2009,24(2):388-392. 被引量:3