Sysmex UF-1000i尿有形成分分析仪对尿液真菌筛查的临床应用评价 被引量:10

Evaluation on the clinical application of Sysmex UF-1000i urine formed element analyzer for screening funguria
暂未订购
导出
摘要 目的评价Sysmex UF-1000i尿有形成分分析仪(简称UF-1000i)在真菌尿筛查中的临床应用价值。方法分别测定UF-1000i检测白念珠菌的精密度、线性范围和携带污染率。选择UF-1000i提示类酵母细胞(YLC)阳性样本200例,阴性样本400例,同时做显微镜检查。结果 UF-1000i检测白念珠菌的精密度为3.83%~4.71%,线性范围为10.0~5 873.1个/μL,线性相关系数(r)为0.999,携带污染率为0.12%。以显微镜检查为金标准,受试者工作特征(ROC)曲线结果显示UF-1000i检测真菌YLC的最佳临界值为18.05个/μL,其检测真菌的敏感性、特异性、准确度、阳性预测值和阴性预测值分别为97.0%、85.6%、87.5%、57.4%和99.3%。结论 UF-1000i可快速提供真菌孢子定量检测结果,辅助诊断尿路真菌感染。 Objective To evaluate the clinical application significance of Sysmex UF-1000i urine formed element analyzer (UF-1000i analyzer) for screening funguria. Methods The precision, linear range and carryover rate of Candida albicans were detected by UF-IO00i analyzer. A total of 200 urine samples which UF-1000i analyzer showed like yeast eell(YLC) positive and 400 YLC negative urine samples were detected by manual microscopy. Results For UF-1000i analyzer detecting Candida albicans, the precision was 3. 83%-4. 71%, the linear range was 10. 0-5 873.1 cells/μL,the correlation coefficient was 0. 999, and the carryover rate was 0.12%. Manual microscopy was as the gold standard, and the result of receiver operating characteristic ( ROC ) curve showed the best cut-off value for UF-1000i analyzer detecting YLC was 18.05 cells/μL. The sensitivity, specificity, accuracy, positive predictive value and negative predictive value of UF-1000i analyzer in the diagnosis of funguria were 97.0% ,85.6% ,87.5% ,57.4% and 99.3%, respectively. Conclusions UF-1000i analyzer can provide quantitative results of fungal spores. It aids to diagnose urinary tract fungal infection.
出处 《检验医学》 CAS 2013年第4期290-292,共3页 Laboratory Medicine
关键词 真菌尿 SYSMEX UF-1000i尿有形成分分析仪 显微镜检查 Funguria Sysmex UF-1000i urine formed element analyzer Microscopy
  • 相关文献

参考文献9

二级参考文献36

共引文献151

同被引文献61

  • 1丛玉隆,马骏龙,张时民,岳秀玲,张清,丁建文.尿液细胞成分定量分析方法学研究[J].中华检验医学杂志,2006,29(3):211-214. 被引量:130
  • 2叶应妩,王毓三,申子瑜.全国临床检验操作规程[M].3版.南京:东南大学出版社,2003:75.
  • 3杨振华,王治国.临床实验室管理[M].北京:人民卫生出版社,2006:1-7.
  • 4王治国.临床检验质量控制技术[M].2版.北京:人民卫生出版社,2010:41-46.
  • 5Beer J H, Vogt A, Neftel K, et al. False positive results for leuco cyles in urine dipstick test with common antibiotics [J]. Bmj, 1996,313(7048) :25.
  • 6Kellogg JA, Manze la JP, Shaffer SN, et al. Clinical relevance of cuhure versos screens for the detection of microbial pathogens in urine specimens[J]. Am J Med, 1987,83(4) :739 -745.
  • 7Males BM, Bartholomew WR, Amsterdam D. Leukocyte esterase nitrile and bioluminescenee assays as urine screens[J]. J Clin Mi erobiol, 1985,22(4) :531- 534.
  • 8Zaman Z, Roggeman S, Verhaegen J. Unsatisfactory performance of flow eytometer UF 100 and urine strips in predicting outcome of urine cultures[J]. J Clin MicroBiol,2001,39(11) :4169- 4171.
  • 9Okada H,Sakai Y, Mfyazaki S,et al. Detection of significant bae teriuria by automated urinalysis using flow cytometry[J]. J Clin Miero Diol,2000,38(8) :2870 -2872.
  • 10Alvarez Barrientos A,Arroyo J,Canton R,et al. Applications of flow eytollletry to e[inical microbiology[J]. Clin Microbiol Rev, 2000,13(2):167- 195.

引证文献10

二级引证文献36

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部