非接触式眼压计与Goldmann压平式眼压计测量值比较研究 被引量:10

Comparison of the intraocular pressure value measured with Goldmann applanation tonometer and non-contact tonometer
原文传递
导出
摘要 目的比较非接触式眼压计(NCT)与Goldmann压平眼压计(GAT)在正常人群及高眼压人群中测量值的差异,并观察比较中央角膜厚度(CCT)对两种眼压计的影响。方法前瞻性临床对比研究。按就诊次序随机选取80例门诊患者共91只眼行CCT、NCT和GAT测量。结果NCT测量值比GAT测量值偏高,且两者测量值差异有统计学意义(F=69.600,P〈0.01),CCT对NCT的影响比GAT偏大,相关系数分别是n=0.742(P〈0.05)和n=0.663(P〈D.05)。结论NCT测量值与GAT测量值有差异,NcT测量平均值比GAT测量平均值偏高,眼压值越高,两者的差异越大;两种眼压计均受CCT的影响,CCT值越大眼压值也越高,且其对NCT的影响大于GAT。 ObJective To compare the difference of Goldmann applanation tonometer (GAT) and non-contact tonometer (NCT) in IOP measurement, and to observe the change of IOP influenced by center corneal thickness (CCT). Methods Central corneal thickness, IOP measured by NCT and Goldmann applanation tonometer were measured on total of 91 eyes of 80 cases. Result There was significant difference between IOP measured by NCT and IOP by Goldmann applanation tonometer (F =69, 600, P 〈0.01), the IOP measured by NCT was higher than GAT, the relative data r1= 0.742 (P 〈0.05) and r2=0.663 (P 〈0.05), so the influence of CCT on NCT was larger than GAT. Conclusions There are differences between IOP measured by NCT and IOP by Goldmann aplana- tion tonometer. IOP measured by NCT is higher than which measured by Goldmann aplanantion tonometer. Both of the two tonometers are influenced by the CCT, and it is larger with the increasing of the CCT. The influence of CCT on the measurements of intraocular pressure with NCT is higher than Golamann aplanation tonometer.
出处 《中国实用眼科杂志》 CSCD 北大核心 2013年第3期270-272,共3页 Chinese Journal of Practical Ophthalmology
关键词 非接触式眼压计 GOLDMANN压平眼压计 中央角膜厚度 Non-contact tonometer (NCT) Goldmann applanation tonometer Center corneal thickness (CCT)
  • 相关文献

参考文献13

二级参考文献43

  • 1段宣初,吴勤,蒋幼芹,卿国平,江冰,石晶明.中央角膜厚度对压平眼压计测量值的影响[J].中国实用眼科杂志,2004,22(10):778-782. 被引量:13
  • 2阎启昌,王欣玲,贺玉华,张劲松.非接触式眼压计与Goldmann压平眼压计测量眼压的比较[J].国际眼科杂志,2006,6(1):114-115. 被引量:10
  • 3胡茂生 刘淳诗.非接触式眼压计与Goldmann氏压平眼压计对比研究[J].中华眼科杂志,1985,:111-111.
  • 4陈惠,眼科研究,1997年,15卷,115页
  • 5李伟力,中华眼科杂志,1997年,33卷,16页
  • 6王浩文,中华眼科杂志,1995年,31卷,172页
  • 7Shieds MB, The non-contact tonometer, Its value and limitations,Surv Ophthalmol, 1980, 24:211.
  • 8Whitacre MM et al. Am J Opthalmol, 1993, 115:592.
  • 9Ehlers N, BransenT, Sperling S, Applanation tonometry and central corneal thickness Acta Ophthalmol, 1975, 53: 34- 43.
  • 10Wolfs RCW et al. Am J Ophthalmol, 1997, 123:767.

共引文献135

同被引文献100

引证文献10

二级引证文献32

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部