非接触式眼压计与Goldmann压平眼压计测量眼压的比较 被引量:11

Comparison between non-contact and Goldmann tonometer in intraocular pressure detection
暂未订购
导出
摘要 目的:比较非接触眼压计(non-contacttonometer,NCT)与Goldmann压平眼压计(GAT)测量眼压的差异,以评价非接触式眼压计在临床青光眼患者及疑似青光眼患者中的的应用价值。方法:志愿者174例348眼分别由专人进行非接触眼压计(NCT)与Goldmann压平眼压计眼压测量。结果:非接触眼压计(NCT)的测量结果低于Gold-mann压平眼压计且差异有显著性(19.6vs23.4,P<0.05),两者相关系数为r=0.783,差异在眼压>30mmHg组更为明显,相关系数为0.334。结论:非接触眼压计(NCT)可以用于临床普查,在青光眼的临床工作中建议使用Goldmann压平眼压计矫正。 AIM: To compare the differences between non-contact (NCT) and Goldmann (GAT) tonometer in intraocular pressure (IOP) detection, so as to evaluate the assessment of the two measurements when used in glaucoma or glaucoma-like patients. METHODS: A total of 174 volunteers (348 eyes) were measured with both NCT and GAT separately. RESULTS: The NCT results were significantly lower than the GAT ones (19.6 vs 23.4, P〈0.05), and the correlation factor r=0.783. Furthermore, there were great differences in the group with IOP〉30mmHg (r=-0.334). CONCLUSION: NCT may be recommended to test the glaucoma-like population while GAT is applied to correct the clinical results in glaucoma patients.
出处 《国际眼科杂志》 CAS 2006年第1期114-115,共2页 International Eye Science
关键词 非接触眼压计(NCT) GOLDMANN压平眼压计 眼内压 non-contact tonometer Goldmann tonometer intraocular pressure
  • 相关文献

参考文献10

  • 1Tonnu PA,Ho T,Newson T,El Sheikh A,Sharma K,White E,Bunce C,Garway-Heath D.The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry,non-contact tonometry,the Tono-Pen XL,and Goldmann applanation tonometry.Br J Ophthalmol,2005;89(7):851-854
  • 2Tonnu PA,Ho T,Sharma K,White E,Bunce C,Garway-Heath D.A comparison of four methods of tonometry:method agreement and interobserver variability.Br J Ophthalmol,2005;89(7):847-850
  • 3Yaoeda K,Shirakashi M,Fukushima A,Funaki S,Funaki H,Ofuchi N,Nakatsue T,Abe H.Measurement of intraocular pressure using the NT-4000:a new non-contact tonometer equipped with pulse synchronous measurement function.J Glaucoma,2005;14(3):201-205
  • 4Lam AK,Chan R,Lam CH.The validity of a new noncontacr tonometer and its comparison with the Goldmann tonometer.Optom Vis Sci,2004;81 (8):601-605
  • 5Ko YC,Liu C J,Hsu WM.Varying effects of corneal thickness on intraocular pressure measurements with different tonometers.Eye,2005;19(3):327-332
  • 6Siganos DS,Papastergiou GI,Moedas C.Assessment of the Pascal dynamic contour tonometer in monitoring intraocular pressure in unoperated eyes and eyes after LASIK.J Cataract Refract Surg,2004;30(4):746-751
  • 7Stabuc Silih M,Hawlina M.Influence of corneal thickness on comparative intraocular pressure measurements with Goldmann and non-contact tonometers in keratoconus.Klin Monatsbl Augenheilkd,2003;220(12):843-847
  • 8Jorge J,Gonzalez-Meijome JM,Diaz-Rey JA,Almeida JB,Ribeiro P,Parafita MA.Clinical performance of non-contact tonometry by Reichert AT550 in glaucomatous patients.Ophthalmic Physiol Opt,2002;22(6):560-564
  • 9Katsushima H,Sone A,Takeda A,Okazaki H,Adachi J,Funahashi K,Maruyama I.Effect of tonometry on a glaucoma population study.Nippon Ganka Gakkai Zasshi,2002;106(3):143-148
  • 10Matsumoto T,Makino H,Uozato H,Saishin M,Miyamoto S.The Influence of Corneal Thickness and Curvature on the Difference Between Intraocular Pressure Measurements Obtained with a Non-contact Tonometer and Those with a Goldmann Applanation Tonometer.Jpn J Ophthalmol,2000;44(6):91

同被引文献43

引证文献11

二级引证文献58

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部