摘要 目的 探讨微电极电刺激 (微刺激 )技术在帕金森病苍白球毁损术的定位作用。方法 采用微电极进行术中电刺激 ,记录不同部位微刺激引起视束及内囊反应的数值。结果 微电极电刺激在靶点区的电刺激电流值在 10 0 0Hz,脉宽 0 2ms视束可达 1~ 10mA ,内囊反应则为 15mA以上。结论 微电极电刺激能精确探及内囊和视束的位置 ,可以有效避免内囊及视束损伤 ,阻止偏瘫和视野缺损的发生。 Objective To probe the role of microelectrode stimulation in pallidotomy.Methods Electric current value of optic tract and internal capsule response to microelectrode stimulation were recorded during the operations.Results In the target territory, current value of stimulation at 1000Hz,0 2ms impulse is 1~10mA for optic tract and more than 15mA for internal capsule.Conclusions Microelectrode stimulation can precisely localize the optic tract and internal capsule in pallidotomies to avoid injury of this two structures.
出处 《中华神经外科杂志》 CSCD 北大核心 2000年第2期75-77,共3页 Chinese Journal of Neurosurgery
关键词 震颤性麻痹 苍白球毁损术 微电极电刺激 定位 Parkinsons disease Microelectrode Pallidotomy
  • 相关文献

参考文献1

二级参考文献2

共引文献38

同被引文献36

  • 1Benabid AL. Deep brain stimulation for Parkinson' s disease[J]. Curr Opin Neurohiol, 2003, 13(6) :696-706.
  • 2Mohr P, Rodriguez M, Slavickova A, et al. The application of vagus nerve stimulation and deep brain stimulation in depression [J]. Neuropsychobiology, 2011,64(3): 170-181.
  • 3Evans AC, Janke AL, Collins DL, et al. Brain templates and atlases[J]. Neuroimage, 2012, 62(2) : 911-922.
  • 4Sumanaweera TS, Adler JR, Napel S, et al. Characterization of spatial distortion in magnetic resonance imaging and its implications for stereotactic surgery[ J]. Neurosurgery, 1994, 35 (4) : 696-704.
  • 5Khan MF, Mewes K, Gross RE, et al. Assessment of brain shift related to deep brain stimulation surgery [ J ]. Stereotact Fnnct Neurosurg, 2008, 86( 1 ) : 44-53.
  • 6Toda H, Sawamoto N, Hanakawa T, et al. A novel composite targeting method using high-field magnetic resonance imaging for subthalamic nucleus deep brain stimulation [ J]. J Neurosurg, 2009, lll (4) : 737-745.
  • 7Pollo C, Villemure JG, Vingerhoets F, et al. Magnetic resonance artifact induced by the electrode Activa 3389: an in vitro and in vivo study[J1. Acta Neurochir (Wien), 2004, 146 (2): 16i-164.
  • 8Karl AS, Deborah R, Buyan-Dent L, et al. Wide-bore 1.5T MRI-guided deep brain stimulation surgery: initial experience and technique comparison [ J]. Clin Neurol Neurosurg, 2014, 127 (9) : 79-85.
  • 9Thani NB, Bala A, Lind CR. Accuracy of postoperative computed tomography and magnetic resonance image fusion for assessing deep brain stimulation electrodes [ J ]Neurosurgery, 2011, 69(1): 207-214.
  • 10Walter U, Kirsch M, Wittstock M, et al. Transcranial sonographic localization of deep brain stimulation electrodes is safe, reliable and predicts clinical outcome [J]. Ultrasound Med Biol, 2011, 37(9) : 1382-1391.

引证文献4

二级引证文献33

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部