期刊文献+

人工耳蜗植入术中蜗内电极X线拍片的临床意义 被引量:8

Plain X-ray for locations of intracochlear electrode arrays during cochlear implantation
原文传递
导出
摘要 目的:探讨耳蜗X线拍片在人工耳蜗植入术中的临床意义。方法:对54例人工耳蜗植入患者术中常规使用可移动C-臂数字X线拍片机拍片,观察耳蜗电极的形态和植入深度,常规行NRT检查。结果:52例蜗内电极均呈自然螺旋状,无扭曲或打折,拍片显示刺激电极全部植入耳蜗内。2例插入蜗内的电极串呈"C"形,提示电极未能全部插入,其中1例与术中所见一致,为电极植入不全;1例为术中未能发现的蜗内电极扭曲打折。结论:人工耳蜗植入术中常规在电极植入后进行耳蜗X线拍片,对了解电极植入部位和植入深度非常有益,特别是对于电极植入有难度的患者,更应及时拍片,可避免术后再次手术。 Objective:To evaluate the position and shape of inserted electrode arrays by plain X-ray during cochlear implantation.Method:Cochlear view of implanted electrode arrays by plain X-ray were observed during operation in 54 patients received cochlear implant.Result:The image in cochlear view could provide the information about position and shape of inserted electrode arrays.The spiral-shape electrode arrays without distortion and folding were found in 52 cases.The complete insertion of introcochlear electrode in these patients were confirmed by surgery.Of 54 cases,the other 2 cases showed improper position and C-shape of intracochlear electrodes,one was partial insertion and the other was bending inside the cochlea.The bending electrode array was reinserted immediately during operation.Conclusion:Routine plain X-ray after inserted electrode during operation was a best way of knowing the shape and depth of the electrode array,especial for the difficult cochlear implantation,and also could avoided anesthetization again.
出处 《临床耳鼻咽喉头颈外科杂志》 CAS CSCD 北大核心 2012年第13期598-600,共3页 Journal of Clinical Otorhinolaryngology Head And Neck Surgery
关键词 人工耳蜗植入 耳蜗电极 X线拍片 cochlear implantation ation intracochlear electrode arrays plain X-ray
  • 相关文献

参考文献11

  • 1李玉洁,张道行.1396例人工耳蜗植入围手术期并发症讨论[J].临床耳鼻咽喉头颈外科杂志,2010,24(10):433-435. 被引量:30
  • 2银力,曹永茂,龙墨,王树峰.X线平片对人工耳蜗植入电极位置判定的意义[J].听力学及言语疾病杂志,2009,17(4):376-379. 被引量:7
  • 3孔维佳,马辉,韩萍,乐建新,熊新高,朱立新,戴陈凯.人工耳蜗植入术后植入电极的影像学检查[J].中华耳鼻咽喉科杂志,2004,39(2):81-84. 被引量:16
  • 4Kim CS,Oh SH,Chang SOet al.Management of complications in cochlear implantation[].Acta Oto laryngologica.2008
  • 5Kubo,T.,Matsuura,S.,Iwaki,T.Complications of cochlear implant surgery[].Op Tech Otorhinolaryngol.2005
  • 6Shpizner,B.A.,Holliday,R.A.,Roland,J.T.Postoperative imaging of the multichannel cochlear implant[].American Journal of Neuroradiology.1995
  • 7Xu J,Xu S A,Cohen L T,et al.Cochlear view: postoperative radiography for cochlear implantation[].American Journal of Otolaryngology.2000
  • 8FRANCIS H W,BUCHMAN C A,VISAYA J M,etal.Surgical factors in pediatric cochlear implantationand their early effects on electrode activation andfunctional outcomes[].Otology and Neurotology.2008
  • 9GROLMAN W,MAAT A,VERDAM F,et al.Spreadof excitation measurements for the detection of elec-trode array foldovers:aprospective study comparing3-dimensional rotational X-ray and intraoperativespread of excitation measurements[].Otology and Neurotology.2008
  • 10BIEDRON S,PRESCHER A,ILGNER J,et al.Theinternal dimensions of the cochlear scalae with specialreference to cochlear electrode insertion trauma[].Otology and Neurotology.2010

二级参考文献18

  • 1杨仕明,张德秀,郗昕,冀飞,韩东一,杨伟炎.多导人工耳蜗植入术后耳蜗X线显像[J].中华耳科学杂志,2003,1(2):46-49. 被引量:9
  • 2张道行,胡宝华,肖玉丽,史波宁.人工耳蜗植入的术前评估与术中处理[J].中华耳鼻咽喉科杂志,2004,39(10):594-597. 被引量:44
  • 3Xu J,Xu SA,Cohen LT,et al.Cochlear views Post-operative radiography for cochlear implantation[J].AM J Otol,2000,21:49.
  • 4Cohen LT,Xu J,Xu SA,et al.Improved and simplified methods for specify the position of the electrode bands of a cochlear implant array[J].AM J Otol,1996,17:859.
  • 5Carelsen B,Grolman W,Tange R,et al.Cochlear implant electrode array insertion monitoring with intra-operative 3D rotational X-ray[J].Clinical Otolaryngology,2007,32:46.
  • 6Viccaro M,De Seta E,Covelli E,et al.Another reason for intra-operative imaging during cochlear implantation[J].The Journal of Laryngology & Otology,2008,122:12.
  • 7Tange RA,Grolman W,Maat A.Intraeochlear misdirected implantation of a cochlear implant[J].Acta Oto-Laryngologica,2006,126:650.
  • 8Eun JS,Sang CK,Jae YC.Cochlear implant electrode misplaced in the carotid canal[J].Arch Otolaryngol Head Neck Surg,2007,133:827.
  • 9Tange RA,Grolman W,Carelsen B.Migration of the ball electrode after cochlear implantation[J].Otology & Neurotology,2007,28:195.
  • 10KEMPF H G, JOHANN K,WEBER B P,et al. Complications of cochlear implant surgery in children[J]. Am J Otol,1997,18(6 Suppl):S62-63.

共引文献48

同被引文献56

  • 1杨仕明,张德秀,郗昕,冀飞,韩东一,杨伟炎.多导人工耳蜗植入术后耳蜗X线显像[J].中华耳科学杂志,2003,1(2):46-49. 被引量:9
  • 2林鹏,马元煦,董红.人工耳蜗植入术的影像学评估[J].中国医学文摘(耳鼻咽喉科学),2009,24(5):255-256. 被引量:1
  • 3余力生,陈峰,郑鸿伟,马鑫.再次人工耳蜗植入术[J].中华耳鼻咽喉科杂志,2004,39(10):598-601. 被引量:7
  • 4黄德健,赵晓埝,邢光前,卜行宽,王德杭.人工耳蜗植入术后的影像检查方法[J].实用放射学杂志,2005,21(6):585-588. 被引量:5
  • 5刘军,戴朴,韩东一.人工耳蜗植入的效果评估[J].中华耳科学杂志,2007,5(1):21-25. 被引量:35
  • 6卜行宽.综述遗传性感音神经性聋[J].国外医学.耳鼻咽喉科分册,1987,11:321-321.
  • 7Berit N Verbrist, JohanHMFrijns, JakobGeleijns, etal,MultisoptionCT. as a valuable tool in the postoperative assessment of the cochlear implant patient[J].AJNR, 2005,26:424-429.
  • 8Marsh MA, Xu J, Blamery PJ, et al. Radiologic evaluation of multichannel lintraco clear implant insertion depth[J]. Am J Otol, 1993, 14: 386-391.
  • 9FRANCIS H W, BUCHMAN C A, VISAYA J M, etal. Surgical factors in pediatric cochlear implantation and theireffects on electrode activationaad functional outcomes [J]. Otol Neurotol, 2008,29:502-508.
  • 10Himi T, Kataura A, Sakata M, etaI. Three-dimensional imaging of the temporal bon using a helical CT scan and Its application in patients with cochlear implantation[J]. ORL, 1996, 58(6):298-300.

引证文献8

二级引证文献27

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部