在线尿素清除率监测与尿素清除指数对血液透析充分性的评价及相关性分析

Evaluation on adequacy of hemodialysis by on-line clearance monitoring and Kt/V of Correlation A-nalysis
原文传递
导出
摘要 目的探讨在线尿素清除率监测(OCM)与尿素清除指数(Kt/V)在评价血液透析充分性中的价值及相关性分析。方法48例在成都军区昆明总医院血透中心行维持性血液透析的终末期肾衰患者,采用德国贝朗Dialog+血透机自带OCM监测仪检测尿素清除指数(Kt/V),同时在透前、透后抽血,应用尿素动力学-室可变容积模型检测Kt/V计算值,并对OCM检测值和Kt/V计算值进行相关性分析;探讨OCM与尿素清除指数(Kt/V)在评价血液透析充分性中的作用。结果OMC与Kt/V比较差异无统计学意义(1.467±0.318v81.629±0.262,P〉0.05),但两者呈正相关(r=0.539,P〈0.01)。结论OCM检测值与应用尿素动力学-室可变容积模型计算的Kt/v值相近,二者关系密切相关,可以作为评价尿毒症血透患者血液透析充分性的一项较可靠指标。 Objective To evaluate the adequacy of hemodialysis by online clearance monitoring (OCM) and Kt/V of correlation analysis. Methods The Kt/V values of 48 maintenance hemodialysis pa- tients in the Department of Nephrology of General Hospital of PLA were accessed by both OCM of Germany BeiLang Dialog + hemodialysis machine and single,pool urea kinetic model. The Kt/V calculation value was also checked by the urea dynamics - room variable volume model. The results generated at different timepoint were compared, including at the real time, before and after appearing through blood by On - line Clearance Monitoring, and correlation analysis of Kt/~ value was performed. Results There was no signif- icant difference on evaluating the level of Kt./v by these two methods. The relationship between the results of the two methods was significantly positive related. ( r = 0. 539, P 〈 0. 01 ). Conclusions During hemodi- alysis, the OCM option provides an accurate tool for continuous on-line monitoring of urea clearance.
出处 《中国医师杂志》 CAS 2012年第2期177-178,181,共3页 Journal of Chinese Physician
关键词 尿素/分析 肾透析 Urea/AN Renal dialysis
  • 相关文献

参考文献5

  • 1Lambie SH, Taal MW, Fluck R J, et al . Analysis of factors associ- ated with variability in haemodialysis adequacy. Nephrol Dial Transplant ,2004,19 (2) :406-412.
  • 2高嘉元,陆任华,朱铭力,张伟明,钱家麒,倪兆慧,严玉澄.在线尿素清除率监测仪检测透析充分性的准确性[J].上海医学,2011,34(3):198-203. 被引量:4
  • 3Prado M , Roa LM, Palma A, et al. Double target comparison of blood-side methods for measuring the hemodialysis dose. Kidney Int, 2005, 68 (6) : 2863 -2876.
  • 4Ficheux A, Gayrard N, Szwarc I, et al. Use of spent dialysate a- nalysis to estimate blood levels of uraemic solutes without blood sampling: urea. Nephrol Dial Transplant . 2010, 25 (3) : 873- 879.
  • 5Kuhlmann U, Goldau R, SamadiN, et al. Accuracy and safety of online clearance monitoring based on conductivity variation. Neph- rol Dial Transplant ,2001,16(5 ) : 1053-1058.

二级参考文献21

  • 1谢华,林洪丽,李平,陈丽光,王松岩,冯青.联机尿素清除率监测和常规采血计算Kt/V对血液透析充分性评估的对照研究[J].中国血液净化,2004,3(11):587-589. 被引量:23
  • 2陆任华,朱铭力,郭永平,倪兆慧,张伟明,钱家麒.在线尿素清除率监测对血液透析充分性的评价[J].上海交通大学学报(医学版),2006,26(5):532-534. 被引量:7
  • 3Carl D E, Feldman G. Estimating dialysis adequacy using ionic dialysance. Ren Fail, 2008, 30. 491-498.
  • 4Daugirdas J T, Schneditz D. Overestimation of hemodialysis dose depends on dialysis efficiency by regional blood flow but not by conventional two pool urea kinetic analysis. ASAIO J, 1995, 41:M719- M724.
  • 5Prado M, Roa L M, Palma A, et al. Double target comparison of blood-side methods for measuring the hemodialysis dose. Kidney Int, 2005, 68. 2863-2876.
  • 6Ficheux A, Gayrard N, Szwarc I, et al. Use of spent dialysate analysis to estimate blood levels of uraemic solutes without blood sampling: urea. Nephrol Dial Transplant. 2010, 25: 873-879.
  • 7Lira P S, Lee H P, Kho B, et al. Evaluation of pre and postdilutional on-line hemodiafihration adequacy by partial dialysate quantification and on-line urea monitor. Blood Purif, 1999, 17: 199-205.
  • 8Daugirdas J T. Second generation logarithmic estimates of single pool variable volume Kt/V: an analysis of error. J Am Soc Nephrol, 1993, 4:1205- 1213.
  • 9Daugirdas J T, Depner T A, Gotch F A, et al. Comparison of methods to predict equilibrated Kt/V in the HEMO Pilot Study. Kidney Int, 1997, 52:1395- 1405.
  • 10Barth R H. Urea modeling and Kt/V: a critical appraisal. Kidney Int Suppl, 1993, 41: S252-S260.

共引文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部