摘要
目的研究Tecnobody康复系统定量评定本体感觉的可靠性。方法健康人9例,其中男4例,女5例,年龄(22.8±0.68)岁,用Tecnobody按系统要求连续评定双足本体感觉3次。用重复测量标准误、简单相关系数、组内相关系数分析每次的耗时(Time)和平均轨迹差(ATE)。结果Time重复测量标准误左右分别为3.07-3.83和6.65-8.44,ATE左右分别为1.33-1.97和1.39-1.91;Time简单相关系数左右分别为0.919和0.62,ATE左右分别为0.808和0.831;Time组内相关系数左右分别为0.893和0.639,ATE左右分别为0.716和0.734。结论该系统具备较好的可靠性,且评估不必重复多次,但其各参数设置及中国人标准参考值范围仍需进一步探究。
Objective To study the test-retest reliability in quantitative measurement of proprioception using Tecnobody rehabilitation system. Methods Nine healthy volunteers [4 males, 5 females, averaged age (22.8 ±0.68) years] participated in three consecutive measurements on both feet by using Tecnobody rehabilitation system for computerized proprioeeptive assessment. Standard elTor of measurement ( SEM ) , correlation coefficient and intra- class correlation coefficient(ICC) obtained from the three consecutive measurements were used to analyze the "Time" used in each measuring session, and the "average track error" (ATE) of the measurements. Results SEM values of "Time" in left foot and right foot were 3.07 - 3.83 and 6.65 - 8.44 respectively. ATE values in left foot and right foot were 1.33 - 1.97 and 1.39 -1.91 respectively. The "Time" and ATE correlation coefficients of left foot / right foot were 0. 919/0.6 and 0. 808/0. 831, l'espectively. The "Time" ICC values were 0. 893/0. 639 for left foot / right foot, respectively ; and the "ATE" ICC values were 0. 716/0. 734 for left foot / right foot, respectively. Conclusion The Tecnobody rehabilitation system provided a fairly good reliability in both " relative" and " absolute" values in quantitative evaluation of proprioception in the feet. These data in a larger amount may be useful for setting up varia bles and the standard values of the local population for reference in proprioceptive rehabilitation.
出处
《中华物理医学与康复杂志》
CAS
CSCD
北大核心
2012年第1期34-37,共4页
Chinese Journal of Physical Medicine and Rehabilitation
关键词
本体感觉
定量评估
可靠性
平均轨迹差
Proprioception
Quantitative assessment
Reliability
Average track error