40台磁共振成像设备状态检测结果分析 被引量:4

Analysis of performance testing on 40 units of MRI system
原文传递
导出
摘要 目的 对我国部分MRI设备进行状态检测,评价我国MRI设备的应用质量现状。方法40台MRI设备(5台超低磁场、15台低磁场、7台中等磁场、13台高磁场)。使用Magphan体模进行状态检测,MRI状态检测参数包括信噪比、均匀度、线性度、层厚、空间分辨率和低对比度分辨率等6项。结果 5台超低磁场成像设备无一合格;9台低磁场设备合格;7台中等磁场设备全部合格;13台高磁场设备全部合格。结论 超低磁场MRI设备性能较差,很难满足临床应用;部分低磁场MRI质量可满足临床应用;中等磁场和高磁场MRI设备质量普遍较好,但也有个别设备的个别参数不合格。 Objective ToassessthequalityofMRIsystemsbytheperformancetesting .Methods The authorsusedaMagphanphantomtodoperformanceteston 4 0setsofMRIsystemincluding 5ultra lowfield , 15lowfield ,7middlefieldand 13highfieldsystems.TheperformanceparametersofMRIsystemweresignal to noiseratio ,uniformity ,linearity ,slicethickness ,spatialresolution ,andlowcontrastresolution .Results All ultra lowfieldstrengthsystemswereunqualified ,9of 15lowfieldsystemswerequalifiedandallof7middlefield systemsand 13highfieldsystemswerequalified .Conclusions Thequalityofmostultra lowfieldsystemsisvery poor.Thequalityoflowfieldsystemsisnotgoodandsomeofthemmeettheclinicalapplication .Thequalityof mostmiddlefieldandhighfieldsystemsissatisfactory ,onlyalittleofthemhassomeunqualifiedparameters.
出处 《中华放射学杂志》 CAS CSCD 北大核心 2000年第1期13-16,共4页 Chinese Journal of Radiology
关键词 质量控制 磁共振成像 状态检测 Qualitycontrol Magneticresonanceimaging Performancetesting
  • 相关文献

参考文献1

二级参考文献1

  • 1全国大型医用设备应用质量检测技术学习班资料汇编(法规、文件等),1997年

共引文献5

同被引文献21

  • 1冯锡钢,温志波,黄凡衡,林意群.脑功能成像时磁共振仪器稳定性的一种简单评价方法[J].第一军医大学学报,2004,24(7):815-817. 被引量:6
  • 2陈春晓,罗立民,陶华,王世杰.MR图像Ghost伪影的校正[J].中国生物医学工程学报,2005,24(3):350-352. 被引量:5
  • 3王凯,张伟伟,唐一源.磁共振脑功能成像信号——噪声及分析处理研究[J].国际生物医学工程杂志,2006,29(5):277-279. 被引量:3
  • 4Price R_K,Axel L,Morgan T,et al.Quality assurance methods and phantoms for magnetic resonance imaging:report of AAPM nuclear magnetic resonance[J].MEDICAL PHYSICS,1990,17(2):287-295.
  • 5Phantom Test Guidance for the ACR MKI Accreditation Program[EB/OL].(2005-01-01)[2010-03-20].http://www.acr.org/accreditation/mri/mri_qc_forms/phantom_test_guidance.aspx.
  • 6胡逸民 张红志 戴建荣 胡逸民 主编.治疗计划的设计与执行[A].胡逸民,主编.肿瘤放射物理学[C].北京:原子能出版社,1999.463-471.
  • 7Khoo VS,Rearmley DP,Finnigan DJ, et al. Magnetic resonance imaging(MRI): considerations and applications in radiotherapy treatment planning. Radiother Oncol, 1997,42:1 - 15.
  • 8Hill DLG, Hawkes DJ,Gleeson MJ,et al. Accurate framlees registration of MR and CT images of the head:Applications in planning surgery and radiation therapy. Radiology, 1994,191:447-454.
  • 9Schad LR, G ademann G, Knopp M, et al. Radiotherapy treatment planning of basal meningiomas: Improved tumor localization by correlation of CT and MR imaging data. Radiother Oncol, 1992,25:56-62.
  • 10倪萍,陈自谦,钱根年,肖慧,许尚文,陈锦华.BOLD-fMRI研究过程若干质量控制因素分析[J].医疗设备信息,2007,22(12):6-9. 被引量:16

引证文献4

二级引证文献19

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部