期刊文献+

Goldmann视野计和Humphrey视野分析仪检测青光眼早期视野变化的比较和评价

A Comparison and an Appraisal of Goldmann Perimeter and the Humphrey Field Analyzer in Testing the Visual Field Changes of Early Glaucoma
暂未订购
导出
摘要 比较和评价了Goldmann视野计(以下简称G)和Humphrey视野分析仪(以下简称H)。G和H发现早期视野缺损的差异无显著性,但检测同一区域视野缺损的强度和范围有统计学差异。G可以检测等视线,H可以测量光阈值,结合使用这两种视野计,有助于发现青光眼更早期视野损害。 A comparable study and an appraisal were made to the Goldmann Perimeter and the Humphrey Field Analyzer. There is no statistic difference between both perimeters in testing early glaucoma field defects, but significant difference in the size and depth of the field defect in the same area. The Goldmann Perimeter could be used to detect the changes of isopters, while the Humphrey Field Analy zer, the light threshold. We prefer to the both perimeters in combination.
出处 《眼科研究》 CSCD 1990年第2期86-89,共4页 Chinese Ophthalmic Research
关键词 青光眼 视野计 视野 光阈值 perimeter, gl aucomatous visual field, light threshold.