颞骨64排CT低剂量扫描及不同重建函数对图像质量的影响 被引量:2

Effects of Different Function Convolution on Image Quality in Low-dose 64-Detector Computed Tomography of Petrous Bone
暂未订购
导出
摘要 目的旨在评价不同重建函数的低剂量64排CT的颞骨图像质量。方法选取临床怀疑颞骨病变而来放射科行64排CT检查的患者63例并随机分为2组:重建函数为FC30组和FC81组,均以30 mA管电流进行扫描,其他扫描条件固定。双盲法测量各组重建图像的噪声、评价图像质量及各解剖结构的显示情况并进行统计学分析。结果随着管电流量的降低,各组图像噪声均增加,同一重建函数的轴位图像噪声大于冠状位图像噪声,FC30图像噪声低于FC81噪声,各组内以及组间差异具有统计学意义(P<0.05),而各组间图像评价和各解剖结构评分却无统计学差异(P>0.05)。结论颞骨采用64排CT低剂量扫描后图像噪声增加,且重建函数和重建方向对图像噪声影响较大,但是图像质量和各解剖结构的显示能够满足诊断要求。 Objective To evaluate the image quality of the petrous bone using low-dose scans of 64-detector computer tomography with different function convolution(FC).Methods Totally 63 cases of the petrous bone examined by Toshiba Aquilion 64-detector CT were collected and divided into 2 groups(FC30 and FC81) randomly.Other scan parameters were fixed.The image noises and quality as well as the visualization of anatomical landmarks of the petrous bone were evaluated by two advanced radiologists with double-blind method.Results Following the reduction of the tube current,the image noise increased in both groups.There were significant statistical differences of image noise not only between the axial images and coronal images of the same FC but also the images of different FC(P 〈 0.05).Moreover,there were no statistical difference in the scores of image and anatomical landmarks between the two groups(P 〉 0.05).Conclusion The use of low-current protocol resulted in a significant increase in the image noise in petrous bone CT,which was also affected by FC,with the satisfied diagnostic results.
出处 《中国医科大学学报》 CAS CSCD 北大核心 2011年第2期149-152,共4页 Journal of China Medical University
基金 辽宁省科学技术计划项目(2008225009-18)
关键词 颞骨 多层 体层摄影术 X线计算机 低剂量 重建函数 petmus bone multislice tomography X-my computer low dose function convolution
  • 相关文献

参考文献9

  • 1Tan JS,Tan KL, Lee JC, et al. Comparison of eye lens dose on neuroimaging protocols between 16- and 64-section multidetector CT: achieving the lowest possible dose [J]. AJNR,2009,30 (2):373- 377.
  • 2Neufang KFR, Zanella FE, Ewen K. Radiation doses to the eye lenses in computed tomography of the orbit and the petrous bone [ J ]. Eur J Radiol, 1987,7(3) :203-205.
  • 3Niu Y, Wang Z, Liu Y, et al. Radiation dose to the lens using different temporal bone CT scanning protocols [ J ]. AJNR, 2010,31 (2) : 226-229.
  • 4Kalra MK, Maher MM,Toth TL, et al. Strategies for CT radiation dose optimization [ J ]. Radiology, 2004,230 (3) : 619-628.
  • 5Lutz J,Jager V, Hempel MJ,et al. Delineation of temporal bone anatomy:feasibility of low-dose 64-row CT in regard to image quality [J]. Eur Radiol, 2007,17 (10) : 2638-2645.
  • 6严超贵,彭谦,范淼,杨智云,陈星,孟悛非.应用64排螺旋CT低剂量技术检查儿童颞骨结构的探讨[J].影像诊断与介入放射学,2008,17(6):245-247. 被引量:10
  • 7亓恒涛,秦维昌,柳澄,王道才,刘传亚,王巍.64层螺旋CT颞骨低剂量扫描的初步研究[J].中华放射学杂志,2007,41(2):117-120. 被引量:30
  • 8Christoph S,Chen XY. Dose optimization in pediatric CT:current technology and future innovations[J ]. Pediatr Radiol,2002,32(10) : 729-734.
  • 9刘景鑫,杨海山.螺旋CT图像噪声影响因素的实验分析[J].中华放射学杂志,2000,34(1):9-12. 被引量:79

二级参考文献13

共引文献116

同被引文献9

引证文献2

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部