期刊文献+

GEMINI GXL16 PET-CT的验收与NEMA标准的应用 被引量:7

Acceptance tests and performance evaluation for GEMINI GXL16 PET-CT
原文传递
导出
摘要 目的 依照美国国家电气制造协会(NEMA)最新版NEMA NU 2-2001标准,对飞利浦公司生产的GEMINI GXL16 PET-CT仪进行验收和性能评估.方法 使用NEMA NU 2-2001标准的测试模型和测试方法,对PET的空间分辨率、能量分辨率、散射分数、计数丢失、灵敏度、PET与CT图像的融合精度等参数进行测试.结果 空间分辨率[(半高宽)mm]轴向为6.06(0cm)和6.44(10 cm),横向为4.65(1cm)和5.98(10 cm);散射分数为34.27%;能量分辨率为18%;中心(0 cm)和偏离中心10cm处的灵敏度分别为8312和8472(计数·s-1·MBq);等效噪声计数率为53.29×103计数/s;清楚分辨直径为10 mm的最小热球;PET与CT图像融合无偏离(模型).结论 该PET-CT仪的PET部分被测试指标均达到出厂要求,通过验收. Objective To evaluate the performance specification of PHILIPS GEMINI GXL16 PETCT through the acceptance tests. Methods Acceptance tests of GEMINI GXL16 PET-CT consist of PET performance measurements and the precise measurement of imaging fusion. The performance measurement of PET was based on NEMA standard (NU 2-2001 version). Results Spatial resolution( FWHM, mm ) was 6.06/6.44 (axial, 2D/3D) and 4.65/5.98 (tansverse, 2D/3D). Scatter fraction (SF) was 34.27%. Energy resolution was 18%. Sensitivity (counts·s-1·MBq-1) was 8312/8472 (center/1 0cm). Noise equivalent count rate was 53.29 kcps. The phantom imaging can recognize small sphere with diameter 10 mm and the imaging fusion was precise. Conclusion The PET measurement results of the PET-CT met the requirement of manufacture's standard, therefore acceptance test was approved.
出处 《国际放射医学核医学杂志》 2010年第4期230-234,共5页 International Journal of Radiation Medicine and Nuclear Medicine
基金 基金项目:河北省普通高等学校强势特色学科肿瘤学组资助(冀教高[2005]52)
关键词 正电子发射断层显像术 体层摄影术 X线计算机 质量控制 Positron-emission tomography Tomography, X-ray computed Quality control
  • 相关文献

参考文献2

二级参考文献6

  • 1冯彦林,贺小红,黄克敏,余丰文,刘德军,袁建伟,袁白虹,苏少弟.Gemini PET/CT系统性能模型测试[J].中华核医学杂志,2005,25(1):54-56. 被引量:9
  • 2NEMA.Performance measurements of positron emission tomographs.NEMA standards publication NU2-2001.National Elecfrical Manufacturers Association.
  • 3IAEA.Quality Control of Nuclear Medicine Instruments 1991.IAEA-TECDOC-602.
  • 4Erdi YE,Nehmeh SA,Mulnix T,et al.PET performance measurements for an LSO-based combined PET/CT scanner using the national electrical manufacturers association NU2-2001 standard.J Nucl Med,2004,45:813-821.
  • 5Watson CC,Casey ME,Eriksson L,et al.NEMA NU 2 performance tests for scanners with intrinsic radioactivity.J Nucl Med,2004,45:822-826.
  • 6陈盛祖,党亚萍,陶军,耿建华.Discovery LS PET/CT的验收及性能评估[J].中华核医学杂志,2004,24(2):117-120. 被引量:19

共引文献19

同被引文献48

  • 1陈香,赵军,管一晖.PET-CT常见伪影[J].国外医学(放射医学核医学分册),2005,29(5):201-205. 被引量:8
  • 2万良荣,黄钢,孙晓光,刘建军,施一平.PET/CT的质量控制和日常维护[J].上海第二医科大学学报,2005,25(12):1303-1305. 被引量:12
  • 3李小华,贺小红,李凡勇,陈萍,陶军,刘炜.Discovery ST PET性能测试与NEMA标准的应用[J].中华核医学杂志,2006,26(2):104-106. 被引量:14
  • 4杨军,沈金毅.PET/CT质量性能检测的探讨[J].医疗卫生装备,2007,28(9):75-77. 被引量:11
  • 5Lewellen TK.Recent developments in PET detector technology[J].Phys Med Biol. 2008,53(17): 287-317.
  • 6NEMA.Performance measurements of positron emission tomographs[S]. Active: NEMA Standards Publication. 2007.
  • 7Conti M. Effect of Randoms on signal to- noise ratio in TOF PET[J].IEEE Transactions on Nuclear Science,2006,53(3) : 1188 1193.
  • 8Conti M. Focus on time-of-flight PET: the benefits of improved time resolution[J].Eur J Nucl Med Mol Imaging, 201 i, 38(6): i 147- l 157.
  • 9Geramifar P, Ay MR. Investigation of time- of-flight benefits in an LYSO-based PET/CT scanner: A Monte Carlo study using GATE[J]. Nuclear Instruments and Methods in Physics Research A, 2011,641 : 121-127.
  • 10Surti S,Karp JS,Muehllehner G. Image quality assessment of LaBr3-based whole body 3D PET scanners, a Monte Calo evaluation[J].Phys Med Biol, 2004,49(19): 4593-4610.

引证文献7