中国实验快堆C31系统HPGe γ谱仪效率刻度的点源模拟 被引量:4

The peak efficiency calibration for the cover gas monitoring system of CEFR's HPGe γ detector
在线阅读 下载PDF
导出
摘要 中国实验快堆燃料破损覆盖气体监测系统γ谱仪采用的是ORTEC公司的HPGe γ谱仪,监测对象是覆盖气体(氩气)。由于很难找到谱仪效率刻度相对应的标准气体体源,对谱仪的效率刻度带来了困难。本文在参考大量文献的基础上利用点源模拟的方法来进行HPGe γ谱仪的全能峰效率刻度,解决了中国实验快堆燃料破损覆盖气体监测系统(C31系统)谱仪效率刻度的困难。 The cover gas monitoring system(C31) of China Experimental Fast Reactor(CEFR) use the ORTEC company's HPGe gamma-ray spectrometer to monitor the radioactive fission gas.It's hard to find a standard volume source to calibrate the peak efficiency for this HPGe detector.This paper uses the calibration method of the peak efficiency by means of point source simulation for this HPGe gamma ray spectrometry.In this way we find a solution to calibrate the peak efficiency for the C31's gamma-ray spectrometer.
出处 《核科学与工程》 CSCD 北大核心 2010年第2期122-126,共5页 Nuclear Science and Engineering
关键词 Γ谱仪 点源模拟 效率刻度 γ spectroscopy point source simulation efficiency calibration
  • 相关文献

参考文献11

  • 1复旦大学 清华大学 北京大学.原子核物理实验方法[M].北京:原子能出版社,1997.93-96,109-117,301-304.
  • 2R G Helmer. Variation of Ge detector efficiency with source diameter and radial source position[J]. Applied Radiation and Isotopes, 1983, 34:1105-1108.
  • 3A P Kushelevsky, Z B Alfassi. Off center gamma ray detection efficiencies of cylindrical single open ended (SOE) Ge(Li) detectors[J]. Nucl Instr and Meth, 1975, 131:93-95.
  • 4Aguiar Julio C. An analytical calculation of the peak efficiency for cylindrical sources perpendicular to the derector axis in gamma ray spectrometry[J]. Applied Radiation and Isotopes, 2008, 66:1123-1127.
  • 5Presler O, Peled O, German U, et al. Off-center efficiency of HPGe detectors [J]. Nuclear Instruments Methods in Physics Research, 2002: 444-450.
  • 6王崇杰,张爱莲,吕建洲.HPGeγ探测器的点源效率函数及其参数确定[J].核技术,2006,29(1):77-80. 被引量:14
  • 7Oren Presler, Uzi German, et al. The validity of the virtual point detector concept for absorbing media[J]. Applied Radiation and Isotopes, 2004: 213-216.
  • 8Mahling S, Orion I, Alfassi Z B. The dependence of the virtual point-detector on the HPGe detector dimensions [J]. Nuclear Instruments and methods in Physics Research A, 2006, 557: 544-553.
  • 9Vargas M J, Timon A F, Diaz N C, et al. Influence of the geometrical characteristics of an HPGe detector[J]. Journal of Radioanalytical and Nuclear Chemistry, 2002, 253(3): 439-443.
  • 10Alfassi Z B, Lavi N , Presler O, et al. HPGe virtual point detector for radioactive disk sources[J]. Applied Radiation and Isotopes, 2007:253-258.