摘要
评估准分子激光原位角膜切削术治疗近视术后使用非接触式压平眼压计测量眼压的准确性。方法:于LASIK术前及术后3个月采用非接触式压平眼压计测量不同程度近视的1005只眼的眼压,并进行比较。同时,将术后眼压变化结果与术后角膜厚度及曲率变化结果进行回归分析。
To study the intraocular pressure results after excimer laser in situ keratomileusis (LASIK) for correction of various degrees of myopia. Methods: Noncontact tonometer was used to study intraocular pressure in 1 005 eyes of spherical equivalent of - 2. 00 D to - 26. 63 D before and 3 months after LASIK surgery. Results: After LASIK, a decrease was observed in the intraocular pressure that was related to the degree of myopia treated. Conclusion: The intraocular pressure measured by noncontact tonometer after LASIK for myopia may be reduced because of corneal thinning and curvature decreasing.
出处
《北京医科大学学报》
CSCD
1999年第2期174-176,共3页
Journal of Peking University(Health Sciences)
关键词
近视
外科手术
眼压
LASIK
Excimer laser in situ keratomileusis Myopia/sum Tonometry