摘要
用KCl涂层的扩散管(KClcoatdenuder)测定了大气中的二价汞(Hg2+)。实验结果表明:扩散管对Hg2+的吸附率>95%;对Hg0的吸附率<5%,实验所用的扩散管对Hg2+最大吸附量不超过30ng,所吸附的Hg2+可以用1mol·L-1HCl定量解吸,冷原子荧光测定Hg,回收率>98%。选择贵州某汞矿区冶炼车间,测定的Hg2+含量为54~73ng·m-3,约占大气总汞浓度的10%~15%。
A KCl coat denuder was employed for the measurement of divalent mercury(Hg2+)in air. Gaseous Hg2+ and Hg0 were collected by the denuder with a efficiency of >95% and <5%,respectively. Maximum Hg load on one denuder should be less than 30 ng and Hg2+ trapped in the denuder can be quantitatively extracted by 1 mol·L-1 HCl and analyzed by cold atomic fluorecence spectrometry. Hg can be recovered from the denuder with a efficiency of >98%. Hg2+ contents of 54~73 ng·m-3 corresponding to about 15% of the total gaseous mercury were determined for air near a Hg _ smelting workshop.
出处
《分析测试学报》
CAS
CSCD
1999年第2期14-16,共3页
Journal of Instrumental Analysis
基金
国家自然科学基金
国家自然科学基金国际合作项目资助