心电电极模拟除颤恢复性能与法拉第电阻关系的研究 被引量:2

A study on the relation between simulated defibrillation recovery and Faradic resistance of ECG electrodes
暂未订购
导出
摘要 心电电极的模拟除颤恢复性能与电极的法拉第电阻有关,法拉第电阻越小,模拟除颤恢复性能越好.研究发现:电极的法拉第电阻与电极芯材料、电极胶的pH值和电极胶中Cl-离子浓度有关. Simulated difibrillation recovery(SDR)of ECG electrodes is related to its Faradic resistance. The smaller Faradic resistance is,the better SDR is. It is found that Faradic resistance of an ECG electrode depends on material of its electrode core, pH value of its electrode gel and concentration of Cl - ion in its electrode gel. The relations among them were studied in detail in this paper.
出处 《浙江大学学报(自然科学版)》 CSCD 1998年第3期330-335,共6页
关键词 心电电极 模拟 除颤恢复性能 法拉第电阻 ECG electrode simulated defibrillation recovery Faradic resistance electrode gel
  • 相关文献

参考文献2

二级参考文献2

  • 1李国栋,S.库卡尔.新型镍电极及其在碱性电池中的应用[J].电池,1991,21(3):5-7. 被引量:8
  • 2A. Delahaye-Vidal,M. Figlarz. Textural and structural studies on nickel hydroxide electrodes. II. Turbostratic nickel (II) hydroxide submitted to electrochemical redox cycling[J] 1987,Journal of Applied Electrochemistry(3):589~599