期刊文献+

溶血实验评价一种新型可切肖性物活性玻璃陶瓷的血液相容性

The Blood Compatibility Assessment of A New Machinable Bioactive Glass- ceramic by Haemolysis Test
暂未订购
导出
摘要 目的评价一种新型可切削生物活性玻璃陶瓷的血液相容性。方法采用直接接触法、浸提液接触法两种溶血实验方法,将该新型可切削生物活性玻璃陶瓷及其浸提液分别在37℃与稀释人血接触,相互作用60min后,离心并测定其上清吸光度,计算材料的溶血率。结果直接接触法和浸提液接触法测得该新型可切削生物活性玻璃陶瓷溶血率分别为1.86%和0.91%。结论该新型可切削生物活性玻璃陶瓷具有良好的血液相容性。 Objective To evaluate the blood compatibility of a new machinable bioactive glass-ceramic.Meth- ods Using direct contact method and extract contact method,we made the new machinable bioactive glass- ceramic or extract be in contact with the diluted blood of healthy volunteer for 60 min,at 37 ~C.Then centrifugate them and measured the absorbance of the supematant.At last calculated the hemolysis rate.Results The hemolysis rate were 1.86% by direct contact method and 0.91% by extract contact method.Conclusion The new machinable bioactive glass-ceramic showed good blood compatibility.
出处 《中国血液流变学杂志》 CAS 2010年第1期51-52,73,共3页 Chinese Journal of Hemorheology
基金 基金项目:辽宁省教育厅资助项目(05L481)容性.
关键词 玻璃陶瓷 溶血实验 血液相容性 glass-ceramic haemolysis test blood compatibility
  • 相关文献

参考文献7

二级参考文献6

  • 1[1]ISO 10993-4:2000 Biological evaluation of medical devices--Part 4: Selection of tests for interactions with blood
  • 2[2]ISO/TR 7405-1984(E):Hemolysis test (in vitro test directly on materials)
  • 3[3]ISO 10993-12:1992 Biological evaluation of medical devices--Part 12:Sample preparation and reference materials
  • 4[4]ISO 10993-5:1992 Biological evaluation of medical devices--Part 5: Tests for cytotoxicity: in vitro methods
  • 5黄占杰,陈小平,朱光明,李彦,刘苏欣.可切削生物活性陶瓷及其在颔面整形外科的应用[J].中国生物医学工程学报,2000,19(1):66-71. 被引量:5
  • 6陈晓峰.生物活性微晶玻璃结构与性能研究[J].山东轻工业学院学报(自然科学版),1991,5(1):1-9. 被引量:3