摘要 [目的]探讨同一系列不同型号凝血分析仪检测凝血酶原时间(PT)、活化部分凝血酶原时间(APTT)和纤维蛋白原(Fg)的结果是否具有可比性。[方法]收集40例不同浓度的患者新鲜血浆,在两个不同的凝血分析仪(CA1500、CA7000)测定PT、APTT和Fg,并对结果进行统计学分析,以医学决定水平处的系统误差来判断检测系统间的结果是否为临床接受。[结果]CA7000的PT测定结果与CA1500比较,差异无显著性意义(P>0.05),相关系数(r)=0.999;CA7000的INR测定结果与CA1500比较,差异无显著性意义(P>0.05),r=0.999;CA7000的APTT测定结果与CA1500比较,差异无显著性意义(P>0.05),r=0.998;CA7000的Fg测定结果与CA1500比较,差异有显著性意义(P<0.01),r=0.985。医学决定水平处的系统误差大部分小于总允许误差的1/2(PT:T±7.5%;APTT:T±7.5%;Fg:T±10%)。[结论]以CA1500作为标准检测系统,CA7000的测定结果具有临床可接受性;应定期对同一实验室不同凝血分析仪进行比对实验,以保证检验结果间的可比性。 [ Objective ] To discuss the comparability of PT, AFFF and Fg results between different coagulation analyzers of the same series. [ Methods ] We collected 40 patients' plasma and detected PT, APTT and Fg by two coagulation analyzers (CA1500, CA7000) respectively. The results were statistically analyzed and clinical acceptability was evaluated by the systemic error at medical determining level. [ Results ] There was no significant difference between PT results detected by CA7000 and those detected by CA1500 (P 〉 0.05) and the correlation coefficient was 0. 999. There was no significant difference between INR results detected by CA7000 and those detected by CA1500 (P 〉 0.05 ) and the correlation coefficient was 0. 999. There was no significant difference between APTr results detected by CAT000 and those detected by CA1500 (P 〉 0.05 ) and the correlation coefficient was 0.998. There was significant difference between Fg results detected by CA7000 and those detected by CA1500 (P 〈0.01 ) and the correlation coefficient was 0.985. Most of the systemic errors were less than half of the limited systemic error ( PT: T ± 7.5 % ; APTT: T ±7.5 % ; Fg: T ± 10% ). [ Conclusion ] The results of CA7000 are acceptable compared to the standard analyzer CA1500. Comparison study must be done periodically among different coagulation analyzers in the same laboratory in order to ensure the comparability of different results.
出处 《大连医科大学学报》 CAS 2010年第1期94-96,共3页 Journal of Dalian Medical University
关键词 凝血分析仪 凝血酶原时间 活化部分凝血酶原时间 纤维蛋白原 比对实验 coagulation analyzer prothrombin time, PT activated partial prothrombin time, APTT fibrinogen, Fg comparison study
  • 相关文献

二级参考文献5

  • 1周新,涂植光.临床生物化学和生物化学检验.北京:人民卫生出版社,2004:177.
  • 2[3]全国卫生专业技术资格考试专家委员会.临床医学检验[M].济南:山东大学出版社,2004.4.
  • 3程烽,中华医学检验杂志,1995年,18卷,111页
  • 4国家质量技术监督局.检测和校准实验室能力的通用要求(GB/T15481-2000 idt ISO/IEC 17025:1999).北京:中国标准出版社,2000.1-15.
  • 5魏吴,丛玉隆.医学实验室质量管理与认可指南.北市:中国计量出版社.2004.72-75.

共引文献194

同被引文献66

引证文献14

二级引证文献49

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部