期刊文献+

六钛酸钾晶须预富集FAAS法测定环境水样中Cr(Ⅲ)

Determination of Cr(Ⅲ)in Environmental Water Sample by FAAS with Potassium Hexaatitanate Whisker Pre-concentration
在线阅读 下载PDF
导出
摘要 研究了六钛酸钾对于Cr(Ⅲ)的吸附性能,提出了用六钛酸钾晶须作为富集剂,预富集、分离水中Cr(Ⅲ)的新方法。在pH值为6.0的条件下,吸附率可达到100%,吸附容量为12.3mg/g。以3mol/L的HNO3作为解脱剂,沸水浴加热40min,可将吸附在六钛酸钾上的Cr(Ⅲ)定量洗脱。火焰原子吸收光谱法(FAAS)的测定结果表明:Cr(Ⅲ)的浓度在0.20~4.0μg/mL范围内线性良好,检出限为3.5ng/mL,相对标准偏差为3.1%。在优化的试验条件下,实测了天然水中Cr(Ⅲ)的含量,加标回收率为94.0%-102.0%。 The adsorption behavior of potassium hexatitanate for Cr (Ⅲ) was studied, and a new method for pre-concentration and separation of Cr(Ⅲ) in water was put forward using its whisker as concentration agent. The adsorption rate is 100% , and the adsorption capacity is 12.3 mg/g at pH 6.0. Cr (Ⅲ) can be eluted from potassium hexatitanate whisker by elutriant ( HNO3 ) of 3 mol/L and heating in boiling water bath for 40 min. The determination results by flame atomic absorption spectrometry (FAAS) show that the linear range of Cr(Ⅲ) is 0. 20 to 4.0 μg/mL. The detection limit is 3.5 ng/mL, and the relative standard deviation (RSD) is 3.1%. Under the optimized test condition, the content of Cr(Ⅲ)in natural water is measured, and the recovery rate with standard addition is 94.0% to 102.0%.
出处 《中国给水排水》 CAS CSCD 北大核心 2008年第22期75-77,81,共4页 China Water & Wastewater
基金 科技部创新基金(05C26213100474) 江苏省研究生培养创新工程(CX07B_176z)
关键词 六钛酸钾晶须 预富集 火焰原子吸收光谱法(FAAS) Cr(Ⅲ) potassium hexatitanate whisker pre-concentration flame atomic absorption spectrometry (FAAS) Cr(Ⅲ)
  • 相关文献

参考文献7

二级参考文献62

共引文献94

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部