期刊文献+

颈椎前路一体化钢板融合器的临床应用

Clinical application of integrated anterior cervical plate cage benezech implant
原文传递
导出
摘要 目的评价颈椎前路一体化钢板融合器治疗颈椎间盘突出症的长期疗效。方法对54例症状性退行性颈椎间盘突出症和急性颈椎间盘突出症患者行前路椎间盘切除减压一体化钢板融合器置入术。术后行日本脊柱学会(JOA)评分,x线片观察融合情况。随访12—79个月,平均53.2个月。结果所有患者术中无并发症发生。随访显示椎间隙高度和脊柱前凸已恢复,无器械断裂、远期不稳定或假关节形成。1例急性椎间盘突出脊髓部分损伤患者和3例症状严重脊髓型椎间盘突出患者症状改善不明显,其余患者治疗效果良好。术后JOA评分为13—17分,平均16.1分,术后改善率为86.7%,融合率100%。结论颈椎前路一体化钢板融合器治疗颈椎间盘突出症有其生物力学优势,是颈椎前路融合固定一项可靠的技术。 Objective To evaluate the long term effect of the integrated anterior cervical plate cage benezech (PCB) implant in treatment of cervical intervertebral disc protrusion. Methods A total of 54 patients with retrogressive and acute cervical intervertebral disc protrusion were treated with anterior decompression and PCB devices of the proper sizes. The results were evaluated by Japanese Orthopaedic Association (JOA) and the fusion rate by X-rays. All patients were followed up for 12-79 months (mean 53.2 months). Results No complication occurred during the operation. The height of intervertebral space and the cervical curvature were regained postoperatively, with no implant breakage, cervical insta- bility or pseudoarticulation. All patients obtained satisfactory outcomes except for one patient with acute spinal cord injury and three with retrogressive cervical intervertebral disc protrusion. The JOA score was 13-17 points (mean 16. 1 points). All patients got fusion, with postoperative improvement rate of 86.7%. Conclusions The integrated anterior cervical PCB implant has the superiority over other im- plants in aspects of clinical application and biomechanics and is a reliable technique for anterior cervical vertebral fusion and stabilization.
出处 《中华创伤杂志》 CAS CSCD 北大核心 2008年第11期868-870,共3页 Chinese Journal of Trauma
关键词 颈椎 椎间盘摘除术 内固定器 脊柱融合术 Cervical vertebrae Intervertebral disk discectomy Internal fixators Spinal fusion
  • 相关文献

参考文献2

二级参考文献10

  • 1严力生,李国栋,章祖成,张继东,钱海平,赵定麟.TFC技术在颈前路手术的应用[J].骨与关节损伤杂志,1996,11(4):210-212. 被引量:7
  • 2Haughton VM.MR imaging of the spine[].Radiology.1988
  • 3George S,Masood S,Peter JH.A new anterior cervical instrumentation system combining an intradiscal cage with an integrated plate[].SPINE.2001
  • 4Clements DH,O leary PF.Anterior cervical discectomy and fusion[].SPINE.1990
  • 5Paramore CG,Dickman CA,Sonntag VKH.Radiographic and clinical follow-up review of Caspar plates in 49 patients[].Journal of Neurosurgery.1996
  • 6Goffin J,Plets C,Van den Bergh R.Anterior cervical fusion and osteosynthetic stabilization according to Caspar: a prospective study of 41 patients with fractures and/or dislocations of the cervical spine[].Neurosurgery.1989
  • 7Kostuik JP,Connolly PJ,Esses SI,et al.Anterior cervical plate fixation with titanium hollow screw plate system[].SPINE.1993
  • 8Lowery GL,McDonough RF.The significance of hardware failure in anterior cervical plate fixation: patients with 2 to 7 year follow-up[].SPINE.1998
  • 9Rizzolo SJ,Piazza MR,Cotler JM,et al.Intervertebral disc injury complicating cervical spine trauma[].SPINE.1991
  • 10赵定麟,赵杰,陈德玉,严力生,李也白.界面固定技术治疗外伤性颈椎不稳[J].中华创伤杂志,1998,14(1):28-30. 被引量:16

共引文献6