摘要 激发光的调制频率和CCD所记录到的光强度是影响荧光寿命成象显微频域法测量和数据处理精度的主要因素.本文采用MonteCarlo方法,以单组分问题为例,研究了在用频域外差法进行荧光寿命成象显微测量中激发光的调制频率以及CCD所记录到的光强度对数据处理精度的影响,给出了当调制频率不变时,在不同荧光强度下,对于不同的荧光寿命范围、荧光寿命计算的统计结果,给出了改进测量精度的方法. In the study of fluorescence lifetime imaging microscopy,modulation frequency of excitationlight and light intensity recorded by CCD are two main factors which affect the precision ofmeasurement and analysis. In this paper, The Monte Carlo method is used to study the two abovementioned factors for the single component problem. The statistical results of lifetime calculation wereobtained for different intensities and lifetime range while the modulation frequency was keptunchanged. Ways to improve the precision of measurement and analysis are given.
出处 《光子学报》 EI CAS CSCD 1997年第9期797-802,共6页 Acta Photonica Sinica
关键词 荧光寿命成象 生物显微术 象增强器 生物光学 Fluorescence lifetime Fluorescence lifetime imaging microscopy Frequency domain MCP intensifier CCD camera
  • 相关文献

参考文献4

  • 1屈军乐,应用激光,1997年
  • 2Wang Xuefeng,Appl Spectrosc,1990年,44卷,1期,25页
  • 3Wang Xuefeng,Appl Spectrosc,1989年,43卷,5期,840页
  • 4方再根,计算机模拟和蒙特卡洛方法,1988年,254页

同被引文献17

  • 1林棋榕,赖克中,李永森,郑蔚,谢树森.计算机图像处理技术在激光-荧光法肺癌定位中的应用[J].激光与光电子学进展,1996,33(4):36-38. 被引量:1
  • 2屈军乐,硕士学位论文,1995年
  • 3李善君,高分子光化学原理及应用,1993年,23页
  • 4Wang Xuefeng,Appl Spectrosc,1990年,44卷,1期,25页
  • 5Wang Suefeng,Appl Spectrosc,1989年,43卷,5期,840页
  • 6屈军乐,生物物理学报
  • 7Wang Xuefeng,Appl Spectrosc,1990年,44卷,1期,25页
  • 8Wang Xuefeng,Appl Spectrosc,1989年,43卷,5期,840页
  • 9方再根,计算机模拟和蒙特卡洛方法,1988年,254页
  • 10屈军乐,生物物理学报,1998年,13卷,4期

引证文献5

二级引证文献27

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部