期刊文献+

应用哈希仪器消解测定水和废水中总磷 被引量:5

The Application of HACH Apparatus to the Total Phosphorus Determination in Water and Waste Water
在线阅读 下载PDF
导出
摘要 建立了利用HACH公司的COD消解一比色管、DRB200 COD加热器及DR890便携式比色计测定水体总磷的方法。确定最佳的操作条件为:过硫酸钾质量分数4%,消解温度120℃,消解时间15min,抗坏血酸溶液的质量分数2%,钼酸盐使用液浓度为储备液1,4。该方法具有较好的精密度和准确度,检出限为0.039nag·L^-1,实际样品分析的相对标准偏差(RSD)在1.5%以下,加标回收率为90.0%.112%。 This paper has brought about a method of determining the total phosphorus in water and waste water by using the COD digest - colorimetrie tube, DRB200 COD heater and DR890 portable colorimeter. The optimal operation condition was verified and the accuracy and precision was satisfactory: detection limit was 0.039 mg·L^- l ; relative standard deviation (RSD) of the specimen was below 1.5 % ; stan- dard recovery of this method was between 90.0% and 112%.
出处 《茂名学院学报》 2008年第1期14-16,共3页 Journal of Maoming College
关键词 总磷 COD消解-比色管 DRB200 COD加热器 DR890便携式比色计 total phosphorus COD digest - colorimetrie tube DRB200 COD heater DR890 portable colorimeter