期刊文献+

Sysmex XE-2100全自动血细胞分析仪对血液病血小板计数的检测价值 被引量:6

暂未订购
导出
摘要 目的:探讨Sysmex XE-2100全自动血细胞分析仪计数血液病血小板的临床应用价值。方法:选用2份低值血小板标本用仪器第4模式检测10次,观察批内精密度。将115例血液病病人血标本用Sysmex XE-2100第3模式电阻法(PLT-I)和第4模式激光法(PLT-O)测定的血小板结果与显微镜计数结果进行相关性分析。结果:2份低值血小板标本用仪器第4模式检测,其批内精密度分别为1.85%和2.03%。Sysmex XE-2100测定115份血液病血小板计数结果与显微镜计数结果相关性良好,激光法相关系数(r)为0.9973,P<0.01;而电阻抗法r为0.9689,P<0.01。结论:Sysmex XE-2100血细胞分析仪激光法(PLT-O)分析血液病血小板计数的性能良好,具有准确度高、重复性好和简便快捷等优点,是常规实验室测定血液病血小板计数较理想的仪器。
出处 《实用医学杂志》 CAS 2007年第7期1072-1074,共3页 The Journal of Practical Medicine
  • 相关文献

参考文献8

  • 1Sysmex公司.仪器的技术指标:Sysmex XE-2100中文操作手册[M].1999:1-14.
  • 2Kunicka J E,Ficher G,Murphy J,et al.Improved platelrt counting using two-dimensional laser light scatter[J].Am J Pathol,2000,114(5):283-289.
  • 3Harrison P,Horton A,Grant D,et al.Immunoplatelet counting:a proposed new reference procedure[J].Br J Heamatol,2000,108 (4):228-235.
  • 4Kunz D.Possibilities and limitations of automated platelet counting procedures in the thrombocytopenic range[J].Seminars in Thrombosis and Hemostasis,2001,27(2):229-235.
  • 5Kunz D,Hoffkes K G,Kunz W S,et al.Standardized flow cytometric method for the accurate detemination of platelet counts in patients with Severe thrombocytopenia[J].Cytometry,2000,42 (5):284-289.
  • 6Harrison P,Horton A,Grant D,et al.Immunoplatelet counting:aproposod new reference procedure[J].Br Jaemato,2000,108(2):283-289.
  • 7林宝顺,郑德柱,刘文星,林祥泉.大血小板检测及临床意义探讨[J].临床检验杂志,2003,21(1):17-18. 被引量:40
  • 8朱忠勇.准确计数血小板方法学研究进展[J].国外医学(临床生物化学与检验学分册),2002,23(3):131-132. 被引量:134

二级参考文献12

  • 1叶应妩 王毓三.全国临床检验操作规程[M].南京:东南大学出版社,1991.70-73.
  • 2川合阳子 清水长子.容易被遗漏的血小板形态的观察[J].日本检查血液学会杂志,2001,2(1):76-76.
  • 3Gill JE, Davis KA, Cowart WJ, et al. A rapid and accurate closed-tube immunoassay for platelets on an automated hematology analyzer[J]. Am J Clin Pathol, 2000,114: 47-56.
  • 4Harrison P, Horton A, Grant D, et al. Immunoplatelet counting: a proposed new reference procedure[J]. Brit J Heamat, 2000,108 : 228-235.
  • 5Kunicka JE, Fischer G, Murphy J, et al. Improved platelet counting using two-dimensional laser light scatter[J]. Am J Clin Pathol, 2000,114 : 283-289.
  • 6England JM, Rowan RM, Bins M, et al. Recommended methods for the visual determination of white cell and platelet count[J]. Available at: 1988/WHO-LAB-88.3.
  • 7Hammerstrom J. Spurious platelet counts in acute leukemia with DIC due to cell fragmentation[J]. Clin Lab Heamat,1992,14: 239-243.
  • 8Matzdorff AC, Kuhnel G, Kemkes-Matthes B, et al.Quantitative assessment of platelets, platelet microparticles,and platelet aggregates with flow cytometry[J]. J Lab Clin Med,1998,131:507-517.
  • 9Dickerhoff R, von Ruecker A. Enumeration of platelets by multiparameter flow cytometry using platelet-specific antibodies and fluorescent reference particles [J]. Clin Lab Haemat, 1995,17:163-172.
  • 10Chapman DH, Hardin J, Miers M, et al. Reduction of the platelet review rate using the two-dimensional platelet method[J]. Am J Clin Pathol, 2001,115: 894-898.

共引文献168

同被引文献30

  • 1肖木洲,郑惠红,梁淑慧,张广聪.静脉血标本静置时间对血细胞分析仪测定结果的影响[J].实用医学杂志,2006,22(15):1822-1823. 被引量:4
  • 2林玉辉,陈美莲,黄仁魏,林桂真,郑永江,黄作平.手工计数与血细胞分析仪测定地中海贫血患者血小板数结果的差异[J].实用医学杂志,2007,23(2):266-267. 被引量:7
  • 3陈梅,方伟祯,傅玉如.Sysmex XE-2100全自动血细胞分析仪分析网织红细胞性能评价[J].实用医学杂志,2007,23(2):268-270. 被引量:20
  • 4邢文理.血细胞检验的质量保证[M].俞善丁.临床基础检验学.北京:人民卫生出版社,2000:104.
  • 5邢文理.红细胞检验[M].俞善丁.临床基础检验学.北京:人民卫生出版社,2000:60-63.
  • 6郭德成.线性相关与回归[M].医学统计学.北京:人民卫生出版社.2004:109-121.
  • 7Pierelli L, Fioravanti D, Patti D,et al. Counting of leukocytes insamples from G-CSF mobilized donors, leukapheresis products,and cord blood: The performances of an analyzer with dedicatedprofiles[ J]. International Journal of Laboratory Hematology,2012,34(2) :154-163.
  • 8Lee HK, Kim SI, ChaeH, et al. Sensitive detection and accuratemonitoring of Plasmodium vivax parasites on routine complete bloodcount using automatic blood cell analyzer ( DxH800 TM ) [ J ].International Journal of Laboratory Hematology ,2012,34(2) :201-207.
  • 9Hedbe^ P, LehtoT. Aging stability of complete blood count andwhite blood cell differential parameters analyzed by Abbott CELL-DYN Sapphire hematology analyzer [ J ] . International Journal ofLaboratory Hematology,2009,31 (1) :87 -96.
  • 10Ang DC,Marinescu L,Kuriyan M.Physician compliance with platelet usagecriteria[J].Arch Pathol Lab Med,2008,132(8):1321-1324.

引证文献6

二级引证文献10