Study on the Long Term Sensing Properties of pCO_2 Sensitive Electrode

Study on the Long Term Sensing Properties of pCO_2 Sensitive Electrode
在线阅读 下载PDF
导出
摘要 In this paper,the pCO_2 sensitive electrode was studied for continuous measurement.The solid electrolyte membrane and gas-permeable membrane were coated on the surface of the pH sensitive electrode which was fabricated based on the SnO_2/ITO glass substrate.According to the experimental results,the pCO_2 sensitive electrode shows the sensitivity of about 35 mV/decade in detection range between 0.1 mmol/L and 50 mmol/L.Moreover,the investigated sensing structure exhibits the convenient and pragmatic properties of the pCO_2 measurement.In summary,the advantages of the pCO_2 sensitive electrode are low-cost and disposable based on the separative structure. In this paper, the pCO2 sensitive electrode was studied for continuous measurement. The solid electrolyte membrane and gas-permeable membrane were coated on the surface of the pH sensitive electrode which was fabricated based on the SnO2/ITO glass substrate. According to the experimental results, the pC O2 sensitive electrode shows the sensitivity of about 35 mV/decade in detection range between 0.1 mmol/L and 50 mmol/L. Moreover, the investigated sensing structure exhibits the convenient and pragmatic properties of the pCO2 measurement. In summary, the advantages of the pCO2 sensitive electrode are low-cost and disposable based on the separative structure.
出处 《稀有金属材料与工程》 SCIE EI CAS CSCD 北大核心 2006年第A03期217-220,共4页 Rare Metal Materials and Engineering
关键词 pCO2 sensitive electrode solid electrolyte membrane gas-permeable membrane SnO2/ITO glass substrate pCO2 sensitive electrode solid electrolyte membrane gas-permeable membrane SnO2/ITO glass substrate
  • 相关文献

参考文献19

  • 1Miller W W, Yafuso M, Yan C F et al, Clinical Chemistry[J], 1987, 33(9): 1538.
  • 2Kohama A, Nakamura Y, Nakamura M et al. Critical Care Medicine[J], 1984, 12(11): 940.
  • 3Weiss I K, Fink S, Harrison R et al. Pediatrics[J], 1999, 103(2): 140.
  • 4Ishiji T, Takahashi K, Kira A. Analytical Chemistry[J], 1993,65(20): 2736.
  • 5Fogt E J, Untereker D F, Norenberg M S et al, Analytical Chemistr[J], 1985, 57(9): 1995.
  • 6Shin J H, Lee H J, Kim C Y et al. Analytical Chemistry[J], 1996, 68(18): 3166.
  • 7Zhang G, Wu X. Optics and Lasers in Engineering[J], 2004, 42(2): 219.
  • 8Cooncy C G, Towe B C. Medical and Biological Engineering and Computing[J], 2004, 42(5): 720.
  • 9Cui G, Lee J S, Kim S J et al. Analytical Chemistry[J], 1998, 123(9): 1859.
  • 10Collison M E, Aebli G V, Petty J et al. Analytical Chemistry [J], 1989, 61(21): 2365.

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部