摘要 目的评价TrapEase腔静脉滤器对肺动脉栓塞的预防效果及使用中的并发症。方法对48例有发生肺栓塞倾向的病人行TrapEase腔静脉滤器置入术。经股静脉或锁骨下静脉或颈静脉途径穿刺插管,具体情况根据血栓位置而定。术后随访3~38个月,平均17个月。检查方法包括腹部X线平片、CT、超声,了解有无发生滤器移位及肺栓塞等并发症。结果48例滤器均置入到预定位置,技术成功率为100%。其中33例放置在肾静脉以下腔静脉段,13例放置在肾静脉以上腔静脉段,2例放置在上腔静脉。术后影像学观察未见滤器移位、断裂及腔静脉损伤。结论对有肺栓塞倾向的上下肢深静脉血栓病人,TrapEase腔静脉滤器是预防肺栓塞发生的有效工具,而且安全可靠。 Objective To evaluate the use of Trap Ease vena cava filter in the prevention of pulmonary embolism and its related complications. Methods A Trap Ease vena cava filter was inserted in 48 patients with tendency of pulmonary embolism by puncture intubation through femoral vein, subclavian vein or jugular vein based on different sites of deep vein thrombosis of lower or upper limb. A follow-up of 3-38 months (mean 17) was made. A plain abdominal radiograph, CT or ultrasound was done to see if there were any complications such as pulmonary embolism or migration of the filters. Results Forty-eight filters were all in right position, the success rate being 100%. Of all the filters, 33 were placed in vena cava under renal vein, 13 above renal vein and two in superior vena cava. No filter migration or fragmentation, no damage of vena cava was noted during follow-up examinations. Conclusion For patients with deep vein thrombosis in upper or lower limps with the tendency of pulmonary embolism, the insertion of Trap Ease filter is an effective and safe means to prevent pulmonary embolism.
出处 《齐鲁医学杂志》 2006年第4期309-311,共3页 Medical Journal of Qilu
关键词 腔静脉滤器 肺栓塞 放射学 介入性 vena cava filter pulmonary embolism radiology, interventional
  • 相关文献

参考文献7

  • 1VAN BEEK E J, TEN CATE J W. The diagnosis of venous thromboembolism : an overvies[M]. // HULL R D, RASKOB G E, PINEO G F. Venous thromboembolism: an evidence based atlas. Amont: Futura Publishing Go,1996:93-99.
  • 2GREEFIELD L J, MCCURDY J R, BROWN P P, et al. A new intraeaval filter permitting continued flow and resolution of embolim[J]. Surgery, 1973,73:599.
  • 3ROEHM J O F, JOHNSRUDE I S, BARTH M H, et al. The bird,s nest inferior vena cave filter progress report[J]. Radiology, 1988, 168:745.
  • 4ATHANSOULIS C A, KAUFMAN J A, HALPERN E F ,et al. Inferior vena caval filter:review of a 26-year single-center clinical experience [J]. Radiology, 2000,216 (10) : 54-66.
  • 5戴汝平.重视肺动脉栓塞的影像学诊断[J].中华放射学杂志,1999,33(5):293-294. 被引量:200
  • 6戴定可,翟仁友,于平.下腔静脉滤器预防肺栓塞的临床应用[J].临床放射学杂志,1999,18(1):48-49. 被引量:41
  • 7RAY C E, KAUFMAN J A. Complications of inferior vena cava filters[J]. Abdomen Image, 1996,21 : 368-374.

二级参考文献3

  • 1Hunsell DM Flower CDR.肺栓塞的影像:应用螺旋CT的新观[J].英国医学杂志:中文版(BMJ),1998,1:6-7.
  • 2Hunsel D M,英国医学杂志,1998年,1卷,6页
  • 3程显声,肺血管疾病学,1993年,179页

共引文献231

同被引文献18

  • 1姚立正,戴真煜,李文会,董从松,张崇国,朱军,朱海东.腔静脉滤器在深静脉血栓治疗中的应用价值及置入体会[J].介入放射学杂志,2007,16(11):737-739. 被引量:14
  • 2Buehler WJ,Gilfrich JV, Wiley RC.Effect of low temperature phase changes on the mechanical properties of alloys near composition TiNi [J].Appl Phys, 1963,34:1475-1477.
  • 3Zadno GR,Duerig TW.Linear and Non-Linear Superelasticity in NiTi[J].MRS Shape Memory Materials, 1989,(9):201-206.
  • 4Cisse O,Savadogo O,Wu L,et al.Effect of surface treatment of NiTi alloy on its corrosion behavior in Hanks' solution[J].J Biomed Mater Res,2002,61 (3):339-345.
  • 5薛淼,陈希贤,李一鸣,等.镍钛记忆合金的基础研究-(I)模拟腐蚀试验[J].口腔医学,1981,1(1):40-43.
  • 6Venugopalan R,Trepanier C.Corrosion of nitinol[A].Russell S,Pelton A.eds.SMST-2000:Proceedings of the International Conference on Shape Memory and Superelastic Technologies[C]. ASM International,2000:261-270.
  • 7Andrea H,Edward W, Andreas A,et al.Comparative MRI compatibility of 316L stainless steel alloy and nickel-titanium alloys stents[J].J Cardiov Magn Reso,2002,4(4):423-430.
  • 8Bard PV.Vena Cava Filters[EB/OL].[2014-02/2016-01-27]. http://www.bardpv.com/vena-cava-filters/.
  • 9An T, Moon E,Bullen Let al.Prevalence and Clinical Consequences of Fracture and Fragment Migration of the Bard G2 Filter Imaging and Clinical Follow-up in 684 Implantations[J].J Vasc lnterv Radiol,2014,25 (6):941-948.
  • 10Stavropoulos SW, Sing RF,Elmasri F,et al.The DENALITrial:An interim analysis of a prospective,multicenter study of the denali retrievable inferior vena cava filter[J].J Vasc Interv Radiol,2014,25( 10 ):1497-1505.

引证文献1

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部