期刊文献+

双压法湿度发生器标准装置测量不确定度评定 被引量:1

Evaluation of Uncertainty on Standard Device of Two Pressure Humidity Generator
在线阅读 下载PDF
导出
摘要 介绍了双压法湿度发生器标准装置的技术指标、测量方法和不确定度评定。测试结果和不确定度分析表明,该发生器在-25-+25℃测量范围内,露点温度测量的不确定度不大于0.04℃。满足我国湿度量值传递的准确度要求。 In the article we introduce the specification of the two pressure humidity generator, meas- urement method, uncertainty component and its evaluation. The measurement results and uncertainty analysis shows that in the dew point range of -25℃- +25 ℃ the uncertainty less than 0.04 ℃. It meets the dissemination requirements of humidity value in this range.
出处 《计测技术》 2006年第B09期90-92,共3页 Metrology & Measurement Technology
基金 科技部科研条件工作项目(JG-2001-35)
关键词 双压法湿度发生器 露点温度 湿度测量 不确定度 two pressure humidity generator dew point temperature humidity measurement uncertainty
  • 相关文献

参考文献3

  • 1Actis A,et al.The use of a humid air generator as a reference method for measuring humidity[A].Proc.of Third International Symposium on Humidity and Moistur[C].Vol.1,London,England:1998.10-19.
  • 2Peter Huang.Determining uncertainties of relative humidity,dew/frost point temperature and mixing ratio in a humidity standard generator[A].Proc.of Third International Symposium on Humidity and Moisture[C].Vol.1,London,England:1998.149-158.
  • 3JJG2046-1990,湿度计量器具检定系统[S].

同被引文献6

引证文献1

二级引证文献10

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部