Reproducibility of measuring retinal nerve fiber layer with scanning laser polarimetry
暂未订购
导出
摘要 目的评价在正常者和青光眼患者中偏振激光扫描仪检查神经纤维层厚度的重复性。方法50例原发性开角型青光眼患者和60例正常者用偏振激光扫描仪行神经纤维层厚度的检查。每例患者同一天内行3次检查,记录下列指标:TSNIT平均值、上方平均值、下方平均值、TSNIT标准差和神经纤维层厚度指示(nerve fi-ber indicator,NFI)值。随机区组方差分析用于检验各次检查之间的差异。计算每例患者三次检查的平均值和标准差,得出所有标准差的平均值及95%可信区间。计算各指标的变异系数(coefficient of variance,CV)及其95%可信区间。结果在正常者和青光眼患者中,各指标三次检查之间差异均无统计学意义(P值均>0.01)。正常者和POAG患者NFI值三次检查的标准差的平均值分别为1.87和3.38。正常者中,TSNIT平均值、上方平均值、下方平均值和TSNIT标准差的平均CV分别为1.67%,2.40%,2.62%和6.16%。POAG患者中,各指标的平均CV分别为2.59%,2.50%,3.98%和6.37%。结论在正常者和青光眼患者中,偏振激光扫描仪的重复性均很好。 Purpose To assess the reproducibility of the measurements with scanning laser polarimetry (SLP) in normal subjects and patients with primary open angle glaucoma (POAG). Methods Totally 60 healthy subjects and 50 patients with POAG were examined with the commercially available SLP GDx with an automated variable corneal compensator. Each subject had 3 scans. Five parameters (TSNIT (temporal, superior, nasal, inferior, temporal) average, superior average, inferior average, TSNIT standard deviation, and nerve fiber indicator (NFI)) were recorded. The measurement reproducibility was assessed. Results The difference between 3 scans for each parameter was not significant (P〉0.01) both in normal subjects and patients. The standard deviation for NFI was 1.87 and 3.38, respectively for normal subjects and patients. Mean coefficients of variation(CV) for TSNIT average, superior average, inferior average, and TSNIT SD were 1.67%, 2.40%, 2.62% and 6.16%, respectively for normal subjects. Mean CV for each parameter were 2. 59%, 2. 50%, 3. 98% and 6.37%, respectively for patients with POAG. Conclusions The reproducibility of measurements with SLP was high both in normal subjects and patients with POAG.
出处 《复旦学报(医学版)》 CAS CSCD 北大核心 2006年第3期408-410,419,共4页 Fudan University Journal of Medical Sciences
关键词 偏振激光扫描仪 重复性 青光眼 视网膜神经纤维层 scanning laser polarimetry reproducibility glaucoma retinal nerve fiber layer
  • 相关文献

参考文献7

  • 1Quigley HA. Neuronal death in glaucoma[J]. Prog Retin Eye Res, 1999,18:39
  • 2Weinreb RN, Dreher AW, Coleman A, et al. Histopathologic validation of Fourier-ellipsometry measurements of retinal nerve fiber layer thickness[J]. Arch Ophthalmol, 1990, 108:557.
  • 3Colen TP, Tjon-Fo-sang M J, Mulder PG,et al. Reproducibility of measurements with the nerve fiber analyzer (NFA/GDx)[J]. J of Glaucoma, 2000, 9(5):363
  • 4Weinreb RN, Bowd C, Zangwill LM. Glaucoma detection using scanning laser polarimetry with variable corneal polarization compensation[J]. Arch Ophthalmol, 2003, 121:218
  • 5Iester M, Mermoud A, Sehnyder C. Frequency Doubling Technique in patients with ocular hypertension and glaueoma[J]. Ophthalmol, 2000, 107:288
  • 6Zhou Q, Weinreb RN. Individualized compensation of anterior segment birefringence during scanning laser polarimetry[J].Invest Ophthalmol Vis Sci, 2002, 43:2221
  • 7Kook MS, Sung K, Park RH, et al. Reproducibility of scanning laser polarimetry (GDx) of peripapillary retinal nerve fiber layer thickness in normal subjects[J]. Graefe's Arch Clin Exp Ophthalmol, 2001, 239:118

同被引文献9

引证文献2

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部